6 Installation of FLUXUS ADM 8027
28
UMFLUXUS_F8V4-6-1EN, 2018-10-10
• Fix the cable gland by screwing the cap nut onto the basic part (see Fig. 6.5).
• Connect the core and the shield correctly to the terminals of the transmitter (see Fig. 6.6, Fig. 6.4 and Tab. 6.1).
6.5.2
Connection of the Extension Cable to the Junction Box
6.5.2.1
Connection without potential separation (standard)
The connection of the extension cable to the junction box without potential separation ensures that transducer, junction
box and transmitter are on the same potential. The extension cable should always be connected in this manner, especially
if there are power current cables nearby the extension cable.
If earthing on the same potential cannot be ensured, see section 6.5.2.2.
• Remove the left blind plug for the connection of extension cable (see Fig. 6.7).
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
Attention!
For good high frequency shielding, it is important to ensure good electrical contact between the exter-
nal shield and the cap nut (and the housing).
Fig. 6.6: Terminals for the connection of the transducers (extension cable)
Tab. 6.1: Terminal assignment (extension cable)
terminal
connection
AV
white or marked cable (core)
AVS
white or marked cable (shield)
ARS
brown cable (shield)
AR
brown cable (core)
Attention!
Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
AV
AV
S
ARS
AR
BVS
BRS
BR
BV
transducers