FLUXUS ADM 8027, F801, ADM 8127B
17 SuperUser-Modus
116
UMFLUXUS_F8V4-6-1EN, 2018-10-10
17
SuperUser-Modus
The SuperUser mode offers the possibility of an advanced analysis of the signal and the measured values as well as the
definition of additional parameters adapted to the measuring point, in order to achieve better measuring values or during
experimental work. Features of the SuperUser mode are:
• Default settings are not complied.
• There are no plausibility during parameter entry.
• There is no check whether the entered parameters are within the limits determined by the laws of physics and technical
data.
• The cut-off flow is not active.
• A value for the number of sound paths has to be entered.
• Some menu items that are not visible in the normal the normal mode are displayed.
17.1
Activation/Deactivation
Enter HotCode
071049
Enter HotCode
071049
again to deactivate the SuperUser mode.
17.2
Transducer Parameters
In the SuperUser mode, the menu item
Transducer Type
will be displayed in the program branch
Parameter
at the
end of the input even if the transducers are detected by the transmitter.
17.3
Defining Flow Parameters
In the SuperUser mode, it is possible to define some flow parameters (profile bounds, correction of the flow velocity) for
the specific application or measuring point.
Attention!
The SuperUser mode is intended for experienced users with advanced application knowledge. The
parameters can affect the normal measuring mode and lead to wrong measuring values or to a fail-
ure of the measurement when a new measuring point is set up.
It is displayed that the SuperUser mode is activated. Press ENTER. The main menu will be
displayed.
It is displayed that the SuperUser mode is deactivated. Press ENTER. The main menu will
be displayed.
Attention!
Some of the defined parameters are still active after the deactivation of the SuperUser mode.
Press ENTER.
or:
Select
Special Version
to enter the transducer parameters. Press ENTER.
If
Special Version
is selected, the transducer parameters have to be entered.
The transducer parameters have to be provided by the transducer manufacturer. Press
ENTER after each input.
Select
Special Funct.
\
SYSTEM settings
\
Measuring
\
Calibration
. Press EN-
TER.
Select the measuring channel for which the flow parameters are to be defined. Press EN-
TER.
SUPERUSER MODE
*IS ACTIVE NOW*
SUPERUSER MODE
IS PASSIVE NOW
Transducer Type
↕
Q2E-314
Transducer Type
↕
Special Version
Transd. Data
1
35.99
Measuring
↕
Calibration
Calibrat. data
↕
for Channel
A: