FLUXUS ADM 8027, F801, ADM 8127B
7 Installation of FLUXUS F801
40
UMFLUXUS_F8V4-6-1EN, 2018-10-10
7.5.2
Connection of the Extension Cable to the Junction Box
7.5.2.1
Connection without potential separation (standard)
The connection of the extension cable to the junction box without potential separation ensures that transducer, junction
box and transmitter are on the same potential. The extension cable should always be connected in this manner especially
if there are power current cables nearby the extension cable.
If earthing on the same potential cannot be ensured, see section 7.5.2.2.
• Remove the left blind plug for the connection of extension cable (see Fig. 7.6).
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
Tab. 7.1: Terminal assignment (extension cable)
terminal
connection
AV
white or marked cable (core)
AVS
white or marked cable (shield)
ARS
brown cable (shield)
AR
brown cable (core)
Attention!
Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Fig. 7.6: Connection of the extension and transducer cable to the junction box JB01
100 mm
20 mm
10
m
m
external shield,
brushed back
extension cable
extension cable
transducer cable
cap nut
compression part
basic part
cable gland
external shield,
brushed back