For the terminal assignment of the transducer cable see Fig. 7.7 and Tab. 7.5.
7.6
Connection of the Power Supply
The external protective earth is connected to the equipotential bonding terminal on the housing of the transmitter (see
Fig. 7.2).
• Remove the cable gland for the connection of the power supply (see Fig. 7.2).
• Prepare the power cable with an M20 cable gland.
• Push the power cable through the cap nut, the compression part and the basic part of the cable gland (see Fig. 7.8).
• Insert the power cable into the housing (see Fig. 7.2).
• Screw the gasket ring side of the basic part into the housing of the transmitter.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the leads to the terminals of the transmitter according to the voltage printed on the nameplate below the termi-
nal strip KL1 (see Fig. 7.3, Fig. 7.9 and Tab. 7.6).
Tab. 7.5: Terminal assignment (transducer cable)
terminal
connection
V
transducer
(core)
VS
transducer
(shield)
RS
transducer
(shield)
R
transducer
(core)
Attention!
Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Attention!
According to IEC 61010-1:2010, a switch has to be provided near the measuring instrument in the
building installation, easily accessible for the user and marked as a disconnection device for the mea-
suring instrument.
If the measuring instrument is used in an explosive atmosphere, the switch should be installed out-
side the explosive atmosphere. If this is not possible, the switch should be installed in the least haz-
ardous area.
Fig. 7.8: Cable gland
Fig. 7.9: Terminals for the connection of the power supply and the outputs
cap nut
compression part
basic part
AV
AV
S
ARS
AR
BVS
BRS
BR
BV
PE
N/L-
L1/L+
PE
N/L-
L1/L+
1 2 3 4
6 7 8
5
9 10 11 12
14 15
13
5-
6+
1-
2+
AV
AR
AV
S
ARS
S S
BV
BR
BVS
BRS
S S
outputs
power
supply
outputs
dividing wall
intrinsically safe
area
not intrinsically safe
area
FLUXUS F801**-A10****-AA, F801**-A10****-AP:
FLUXUS F801C24, F801**-A10****-FF:
power
supply