FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
88
of
170
Peak Hold
APPLICATION: Up to 10 broken points of the material are detected at compression and tension testing
machines. All peak values of testing can be shown on the display automatically after the testing or by pressing
programmed key to the LAST test.
RELATED PARAMETERS: Sub-blocks 31- and 32-. Parameters 241, 242 or 243.
The tension testing process is shown in the diagram below.
Figure 10.7 - Timing diagram of tension test process.
10.4.1 Product entries
Product entries at peak hold are;
Entry type
Start to
follow
Reduction
ratio for
Medium peak
detecting
Test
ending
reduction
ratio.
Stop
testing
Alarm
Unloaded
scale limit
Value
FOLLOW
MID
DECR
END
DECR
STOP
ALARM
EMPTY
Deviation
FOLLOW
MID
DECR
END
DECR
STOP
ALARM
EMPTY
Percent %
FOLLOW
MID -%
END -%
STOP
ALARM
EMPTY
FOLLOW:
The tracing of the force is started from this value.
MID DEC:
Minimum force decrease to detect medium peaks.
END DECR:
Force decrease to finish the test.
STOP:
Stop testing if the force is bigger,
ALARM :
Produce alarm signal to warn the operator and/or to stop testing.
EMPTY :
Test starts automatically or manually if the force is bigger than.
The MID DECR and END DECR entries can be setup as value or percent (refer to par. 312 ) and others are
always values.