Kabelanschlusskasten Typ KAEX         Technisches Handbuch, Rev. 1.02   September 2016 
Junction Box              Type KAEX       Technical Manual 

 

Seite / Page 6 / 8  

RIGHTS

 

AND

 

LIABILITIES 

 

All rights reserved. 

 

No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, 
mechanical, photocopying, recording, or otherwise, without the prior written permission of Flintec GmbH 

 

No patent liability is assumed with respect to the use of the information contained herein. While every precaution has been 
taken in the preparation of this book, FLINTEC assumes no responsibility for errors or omissions. Neither is any liability 
assumed for damages resulting from the use of the information contained herein. 

 

The information herein is believed to be both accurate and reliable. FLINTEC, however, would be obliged to be informed if 
any errors occur. FLINTEC cannot accept any liability for direct or indirect damages resulting from the use of this manual. 

 

FLINTEC reserves the right to revise this manual and alter its content without notification at any time. 

 

Neither FLINTEC nor its affiliates shall be liable to the purchaser of this product or third parties for damages, losses, costs, 
or expenses incurred by purchaser or third parties as a result of: accident, misuse, or abuse of this product or 
unauthorized modifications, repairs, or alterations to this product, or failure to strictly comply with FLINTEC operating and 
maintenance instructions. 

 

FLINTEC shall not be liable against any damages or problems arising from the use of any options or any consumable 
products other than those designated as Original FLINTEC Products. 

 

NOTICE: The contents of this manual are subject to change without notice. 

 

Copyright © 2007 by Flintec GmbH, 74909 Meckesheim, Bemannsbruch 9, Germany 

 

S

AFETY 

I

NSTRUCTIONS

 

 

CAUTION

 

READ this manual BEFORE operating or servicing this equipment. FOLLOW these 

instructions carefully. SAVE this manual for future reference. DO NOT allow untrained personnel to 
operate, clean, inspect, maintain, service, or tamper with this equipment. ALWAYS DISCONNECT this 
equipment from the power source before cleaning or performing maintenance. CALL FLINTEC 
ENGINEERING for parts, information, and service. 

 

 

WARNING 

ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS EQUIPMENT. 

EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE 
WITH POWER ON. FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM. 

 

 

WARNING 

FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT TO 

PROPERLY GROUNDED OUTLET ONLY. DO NOT REMOVE THE GROUND PRONG. 

 

 

WARNING 

DISCONNECT ALL POWER TO THIS UNIT BEFORE REMOVING THE FUSE OR 

SERVICING. 

 

 

WARNING 

BEFORE CONNECTING/DISCONNECTING ANY INTERNAL ELECTRONIC  

COMPONENTS OR INTERCONNECTING WIRING BETWEEN ELECTRONIC EQUIPMENT ALWAYS 
REMOVE POWER AND WAIT AT LEAST THIRTY (30) SECONDS BEFORE ANY CONNECTIONS OR 
DISCONNECTIONS ARE MADE. FAILURE TO OBSERVE THESE PRECAUTIONS COULD RESULT 
IN DAMAGE TO OR DESTRUCTION OF THE EQUIPMENT OR BODILY HARM. 

 

 

CAUTION 

OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE 

DEVICES. 

 

 

I

NTRODUCTION AND TECHNICAL 

D

ATA

 

The painted Aluminium junction box is designed for the parallel connection of 4 load cells. 2 versions are 
available. 

 

Type 

No. of load cells 

Housing size 

Inputs 

Output 

Version  Designation 

KAEX-4  up to 4 

80 x 250 x 52 mm  4x M16 

1x M20 

EEx ia IIC T6 

EEx e IIC T6 

 

The junction box type KAEX can be connected to the instrumentation with a shielded 6-wire signal cable. 
The corner correction is done with resistors. 

 

Housing material: 

Aluminium 

Protection class: 

IP66 

Cable connection: 

With screw terminals 

Corner correction: 

By exchangeable resistors 

 

M

ECHANICAL 

I

NSTALLATION

 

Look for a mounting location which is more or less dry and protected from environmental stress. 

 

Summary of Contents for KAEX-4

Page 1: ...Kabel Anschlusskasten Junction Box Typ KAEX 4 Type KAEX 4 Technisches Handbuch Technical Manual Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www flintec com...

Page 2: ...Mechanische Installation 4 Elektrische Anschl sse 4 Anschluss der W gezellenkabel 4 Anschluss des Ausgangskabels 4 Eckenabgleich bei Waagen mit FLINTEC W gezellen 5 Table of contents RIGHTS AND LIABIL...

Page 3: ...dbuchs ohne vorherige Ank ndigung sind vorbehalten Copyright 2007 der Flintec GmbH 74909 Meckesheim Bemannsbruch 9 Germany SICHERHEITSHINWEISE VORSICHT LESEN Sie diese Handbuch VOR dem Betrieb oder de...

Page 4: ...lb Kabelschirm 1 2 23 24 rot Signal Ausgang 15 16 17 18 weiss Signal Ausgang 11 12 13 14 schwarz Speisung Excitation Eingang 19 20 21 22 ggfs braun R ckf hrung Sense gr n Speisung Excitation Eingang 7...

Page 5: ...r die nun anschlie ende Korrektur Dementsprechend wird die Differenz der anderen Ecken in Bezug auf diese Basisecke errechnet 3 Berechnung des Korrekturwiderstandes wie folgt Korrekturwiderstand in Ab...

Page 6: ...any SAFETY INSTRUCTIONS CAUTION READ this manual BEFORE operating or servicing this equipment FOLLOW these instructions carefully SAVE this manual for future reference DO NOT allow untrained personnel...

Page 7: ...Output 15 16 17 18 white Signal Output 11 12 13 14 black Excitation Input 19 20 21 22 if applicable brown Sense green Excitation Input 7 8 9 10 if applicable blue Sense if load cell is equiped with 6...

Page 8: ...rner correction The differences of the other corners are calculated with reference to this basic corner 8 Calculate the correction resistance as follows Correction resistance Deviation in kg X Input r...

Reviews: