background image

Kabelanschlusskasten Typ KAEX         Technisches Handbuch, Rev. 1.02   September 2016 
Junction Box              Type KAEX       Technical Manual 

 

Seite / Page 8 / 8  

 

Figure 2: Wiring 
 

C

ORNER 

C

ORRECTION AT 

S

CALES WITH 

F

LINTEC 

L

OAD 

C

ELLS

 

Flintec load cells are manufactured with rather tight tolerances, so in most cases an additional corner 
correction is not required. The best conditions are achieved if you use load cells of the same class 
(Designation is done with capital letters A to I on the load cell package besides the type label). 
Hint: Corner errors can have a mechanical background, e.g. sloped mounting surface of the load cell. 

 

Procedure: 
6.  Get the display value for each corner. Use the highest possible display resolution (e.g. factor 10 or higher) 

or, if this is not possible, measure the the digital weight increment using corresponding weights. 

7.  The corner with the lowest display value is the starting point for the corner correction. The differences of 

the other corners are calculated with reference to this “basic corner”. 

8.  Calculate the correction resistance as follows: 

 

Correction resistance 

[

]  =

 

Deviation in [kg] 

 X  Input resistance of the load cell [

] * 

Test load in [kg] 

 

* 1100 

 input resistance for: 

BK2, SB4, SB5, SB6, SB14, SLB, ZLB, UB1, UB5, UB6, PB, RC3 

    400 

 input resistance for: 

RC1, SB2 

 

Example 1:

  1100 

 load cell 

0.1 kg corner error with 500 kg test load 

0.1 kg 

1100 

 

0.22 

 

500 kg 

 

Example 2:

  400 

 load cell 

10 kg corner error with 5000 kg test load 

10 kg 

400 

 

0.8 

 

5000 kg 

 

 

9.  Install the correction resistor into the junction box for the corresponding load cell excitation, e.g. for load 

cell no. 1 the +excitation wire has to be changed from terminal no. 7 to terminal no. 3 and afterwards the 
correction resistor has to be inserted between terminal no. 3 and terminal no. 7. 

10.  Check the corners again. If required repeat the procedure. 
 
Hint: 50 ppm resistors for corner correction are available as a set of 14 values from 0.22 

 to 4.7 

 (10 pcs. 

for each value; article no. 5200-030). 
 
Check the cover sealing and install the cover of the box. 
 

Summary of Contents for KAEX-4

Page 1: ...Kabel Anschlusskasten Junction Box Typ KAEX 4 Type KAEX 4 Technisches Handbuch Technical Manual Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www flintec com...

Page 2: ...Mechanische Installation 4 Elektrische Anschl sse 4 Anschluss der W gezellenkabel 4 Anschluss des Ausgangskabels 4 Eckenabgleich bei Waagen mit FLINTEC W gezellen 5 Table of contents RIGHTS AND LIABIL...

Page 3: ...dbuchs ohne vorherige Ank ndigung sind vorbehalten Copyright 2007 der Flintec GmbH 74909 Meckesheim Bemannsbruch 9 Germany SICHERHEITSHINWEISE VORSICHT LESEN Sie diese Handbuch VOR dem Betrieb oder de...

Page 4: ...lb Kabelschirm 1 2 23 24 rot Signal Ausgang 15 16 17 18 weiss Signal Ausgang 11 12 13 14 schwarz Speisung Excitation Eingang 19 20 21 22 ggfs braun R ckf hrung Sense gr n Speisung Excitation Eingang 7...

Page 5: ...r die nun anschlie ende Korrektur Dementsprechend wird die Differenz der anderen Ecken in Bezug auf diese Basisecke errechnet 3 Berechnung des Korrekturwiderstandes wie folgt Korrekturwiderstand in Ab...

Page 6: ...any SAFETY INSTRUCTIONS CAUTION READ this manual BEFORE operating or servicing this equipment FOLLOW these instructions carefully SAVE this manual for future reference DO NOT allow untrained personnel...

Page 7: ...Output 15 16 17 18 white Signal Output 11 12 13 14 black Excitation Input 19 20 21 22 if applicable brown Sense green Excitation Input 7 8 9 10 if applicable blue Sense if load cell is equiped with 6...

Page 8: ...rner correction The differences of the other corners are calculated with reference to this basic corner 8 Calculate the correction resistance as follows Correction resistance Deviation in kg X Input r...

Reviews: