6.9 Non-Contact Voltage Detector
Warning: For safety, always test the voltage detector on a known live circuit before
using the voltage detector on unknown voltage potentials.
1.
Set switch to the NCV
position. See Figure 6-3.
2.
Be sure to remove the test leads from the meter.
3.
Short press the
RANGE button to toggle the Hi (20~1000V) and Lo (80~1000V)
sensitivity modes.
4.
Position the top of the meter close to a source of voltage.
5.
When the meter detects a voltage, it beeps and displays a blue bar animation. The
tone rate and the bar number will increase relative to the detected voltage.
Fig. 6-3 Non-Contact Voltage Detector