7.17
Changing camera lenses
Applicability
: Camera models with an exchangeable lens.
Note
If the new lens has not been used with the camera before, the lens–camera com-
bination must be calibrated after the lens has been mounted. See section 7.18
ing the lens–camera combination
for information on how to do this.
Note
Do not touch the lens surface when you change lenses. If this happens, clean the
lens according to the instructions in 7.12.2
Follow this procedure:
1. Take a firm grip around the inner ring of the lens. Rotate the inner ring 30° counter-
clockwise until it stops.
#T810583; r. AA/75420/75423; en-US