Setup — Screening station (Auto and Button mode)
2. Screening position.
It is important to clearly mark the screening position (where the person should stand
for the screening). The distance from the camera lens to the center of the screening
position shall be 160 cm (5.2 ft.).
3. Queueing system, e.g. barriers and arrow floor stickers.
The queueing system will guide the visitors into the screening station, to the correct
screening position, and out of the screening station. A recommended standard is to
keep the queue at least 90 cm (36 in.) wide. To make sure people do not cut the cor-
ner and end up too close to the camera, about 45 cm (18 in.) should be added to the
barrier leading in to the screening station.
4. Optional: Backdrop.
A backdrop may be needed to prevent background disturbances from reaching the
camera, such as passing or queueing people.
Additional, not in the figure:
• Bluetooth button (Remote operation button).
Used in Button mode to trigger the screening.
• Optional: Operator display.
Secondary display placed near the operator for monitoring.
• Optional: Information roll-up.
Roll-up that informs people what the screening is about and why they need to do it.
• Optional: Instruction roll-up.
Roll-up that tells people to remove eyeglasses before they are screened.
Note
Removing eyeglasses is crucial for a correct screening result.
10.2
Equipment, materials, and tools
This chapter describes the equipment and material needed to set up a screening station
for Auto or Button mode screening.
10.2.1
Camera accessories
The following accessories, included in the camera package, are needed:
• USB Type-C to USB Type-C cable
• HDMI adapter (USB Type-C to HDMI and PD adapter)
• Power adapter (Power supply, 15 W/3 A)
• Remote operation button (needed for Button mode)
10.2.2
Display
The following display equipment is needed:
• Visitor display.
• HDMI cable.
• Power cable.
• Optional: Operator display.
• Optional: HDMI splitter for connection of two displays.
10.2.3
Equipment and supporting material
The following equipment and supporting material are needed:
• Mounting equipment for display and camera; e.g. stand, wall mount, tripod, bar table.
• Marker to indicate the screening position; e.g. floor sticker.
• Queueing system; e.g. barriers, arrow stickers.
• Optional: Backdrop, preventing background disturbances.
• Optional: Information materials; e.g. roll-ups.
Note
For available roll-ups and other supporting materials from FLIR Systems, go to
https://www.flir.com/
.
#T810583; r. AA/75420/75423; en-US