Setup — Screening station (Auto and Button mode)
10.3.6
Define the screening position
1. Make a straight line from the camera towards the screening position - use a masking
tape.
2. Make a mark on the tape at 160 cm (5.2 ft.) from the camera lens, indicating the
screening position.
10.3.7
Adjust the angle of the camera
1. Place a 170 cm (5.6 ft.) tall person at the screening position.
2. Tilt the camera so that it aims at the face of the person.
3. Look at the image on the display and adjust the angle of the camera until the eyes of
the person are inside the box.
10.3.8
Test the setup (Auto mode)
To test the setup for Auto mode screening, do the following:
1. Enter initiation mode by first selecting another Recording mode and then selecting
the Screening mode again.
To change the Recording mode, select
(
Settings
) >
Recording mode
.
2. In the
Screening
dialog box, select
Auto
.
3. The camera must first record an initial sample for the reference temperature
calculation.
Place a person at the marking on the tape that indicates the screening position. The
camera graphics may indicate that the person needs to move closer to the camera,
which is OK to do.
4. Make sure the person looks into the camera and stands still until a screening result is
displayed.
5. Check that the camera focus is correct and that the measurement was done at the
tear duct. If not, deactivate and activate the Screening mode again and restart the
initiation.
#T810583; r. AA/75420/75423; en-US
44