2. In the
Screening
dialog box, do the following:
• Select the option
Operator
.
• Select if there should be a beep or no sound when the system detects an elevated
temperature.
• Select the allowed deviation from the sampled average. It is recommended to start
with the default value 1.0
℃
/1.8
℉
.
13.5
Record reference samples
Before you can start the screening, you must record reference samples. These are used
to calculate the reference temperature average.
You record the reference samples by screening a healthy person. Make sure this person
represents normal people who will pass the screening (e.g. coming directly from out-
doors, since this affects the skin temperature).
1. Aim the camera toward a healthy person. Make sure the person is in a good position
for the screening.
Note
• Removing eyeglasses is crucial for a correct screening result.
• The eyes must not be covered.
2. Adjust the camera focus by rotating the focus ring or by pushing the autofocus button.
Note
It is important to adjust the focus correctly. Incorrect focus adjustment affects
the temperature measurement.
3. To record a sample, tap
Sample
on the screen or push the P button.
Note
Push the P button shortly. If you push and hold the P button, the sampled
average will be reset.
#T810583; r. AA/75420/75423; en-US
63