Articulation Setup and Orientation
The articulation direction and orientation may change if the probe is twisted and coiled during use or
when stored. To re-orient the probe:
1.
Place the extended probe on a flat surface
2.
Set the tension “F” knob to the left or released position.
3.
Set the articulation knob to the center position with the R
Δ
and
Δ
L labels
in the horizontal position.
4.
Manually adjust the probe and camera to the zero degree position.
5.
Rotate the articulation knob clockwise for a 90° bend.
6.
If necessary, manually adjust the probe so that the camera lays flat on the
surface with a 90° articulation
7.
Rotate the articulation knob counter-clockwise and confirm proper
orientation.
Specifications
Resolution
640 x 480 pixels
Length
1m (3.3’)
Outer diameter (O.D.)
6mm (0.02”)
Focus distance
2 to 6cm (0.79 to 2.36”)
Operating Temperature
-20 to 60°C (-4 to 160°F)
Storage Temperature
-40 to 80°C (-40 to 176°F)
Warranty
1 year
Copyright
©
2011 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.
Revision 2.1 11/11
RΔ
ΔL