background image

18                                  MG302-EU-EN V2.0   4/11 

 

Meg OHMS 

Terminal Voltage 

Range 

Resolution  Accuracy 

Test current  Short circuit 

current 

125V (0%~+10%)  0.125~4.000 M

  0.001M

 

+(2%+10) 

1mA @ load 

125k

 

1mA 

4.001~40.00 M

  0.01M

 

+(2%+10) 

40.01~400.0 M

  0.1M

 

+(4%+5) 

400.1~4000 M

 

1M

 

+(5%+5) 

250V (0%~+10%)  0.250~4.000 M

  0.001M

 

+(2%+10) 

1mA @ load 

250k

 

1mA 

4.001~40.00 M

  0.01M

 

+(2%+10) 

40.01~400.0 M

  0.1M

 

+(3%+5) 

400.1~4000 M

 

1M

 

+(4%+5) 

500V (0%~+10%)   0.500~4.000 M

  0.001M

 

+(2%+10) 

1mA @ load 

500k

 

1mA 

4.001~40.00 M

  0.01M

 

+(2%+10) 

40.01~400.0 M

  0.1M

 

+(2%+5) 

400.1~4000 M

 

1M

 

+(4%+5) 

1000V 

(0%~+10%) 

1.000~4.000 M

  0.001M

 

+(3%+10) 

1mA @ load 

1M

 

1mA 

4.001~40.00 M

  0.01M

 

+(2%+10) 

40.01~400.0 M

  0.1M

 

+(2%+5) 

400.1~4000 M

 

1M

 

+(4%+5) 

 

 
Note:
 Accuracy specifications consist of two elements: 

 (% reading) – This is the accuracy of the measurement circuit. 

 (+ digits) – This is the accuracy of the analog to digital converter. 

 

Summary of Contents for Extech Instruments MG302

Page 1: ...User s Guide Wireless TRMS Multimeter and Insulation Tester Model MG302...

Page 2: ...nd exceeds in this case 1000 VAC or VDC This symbol adjacent to one or more terminals identifies them as being associated with ranges that may in normal use be subjected to particularly hazardous volt...

Page 3: ...IEC 61010 2 USE EXTREME CAUTION when working with high voltages 3 DO NOT measure voltage if the voltage on the COM input jack exceeds 1000V above earth ground 4 NEVER connect the meter leads across a...

Page 4: ...unciators Continuity Diode test Battery status n nano 10 9 capacitance micro 10 6 amps cap m milli 10 3 volts amps A Amps k kilo 103 ohms F Farads capacitance M mega 106 ohms RF transmitter active Ohm...

Page 5: ...be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets As a result the reading may show 0 volts when the outlet actually...

Page 6: ...lack test lead banana plug into the negative COM jack 2 For current measurements up to 4000 A DC set the function switch to the A position and insert the red test lead banana plug into the A mA jack 3...

Page 7: ...Touch the black test probe tip to the neutral side of the circuit Touch the red test probe tip to the hot side of the circuit 9 Apply power to the circuit 10 Read the current in the display 11 In the...

Page 8: ...M jack and the red test lead banana plug into the positive V jack 3 Press the MODE button to indicate and V on the display 4 Touch the test probes to the diode under test Forward voltage will typicall...

Page 9: ...es 5 Read the temperature in the display Note The temperature probe is fitted with a type K mini connector A mini connector to banana connector adaptor is supplied for connection to the input banana j...

Page 10: ...ss the TEST button to start a hands free test Note If the circuit under test is live and has a voltage potential AC DC over 30V the meter it will not test the display will show 30V the symbol will fla...

Page 11: ...appear The right most auxiliary display will show and hold the minimum reading updating only when a new MIN occurs 2 To exit MAX MIN mode press EXIT RELATIVE MODE The relative measurement mode displa...

Page 12: ...the desired sample rate 0 to 255 seconds 4 Set the sample rate to 0000 S for manual recording In this mode each press of the STORE button will save one measurement reading 5 Set the sample rate from...

Page 13: ...tions SETUP 1 Press and Hold the RANGE SETUP button for two seconds to enter the SET function The first of five settable functions will appear 2 Press the RANGE button to step through the functions a...

Page 14: ...e old or weak batteries so they do not leak and damage the unit 6 IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME the batteries should be removed to prevent damage to the unit BATTERY INSTALLAT...

Page 15: ...cessible 3 To replace the 10A fuse remove the six screws securing the rear cover and remove the rear cover 4 Gently remove the old fuses and install the new fuses into the holders 5 Always use a fuse...

Page 16: ...1000V 1V All AC voltage ranges are specified from 5 of range to 100 of range DC Current 400 A 0 01 A 1 0 reading 3 digits 4000 A 0 1 A 40mA 0 001mA 400mA 0 01mA 10A 0 001A 20A 30 sec max with reduced...

Page 17: ...z 0 0001MHz 40MHz 0 001MHz 100MHz 0 01MHz Not specified Sensitivity 0 8V rms min 20 to 80 duty cycle and 100kHz 5Vrms min 20 to 80 duty cycle and 100kHz Frequency electrical 40 00HZ 4KHz 0 01HZ to 0 0...

Page 18: ...mA 4 001 40 00 M 0 01M 2 10 40 01 400 0 M 0 1M 3 5 400 1 4000 M 1M 4 5 500V 0 10 0 500 4 000 M 0 001M 2 10 1mA load 500k 1mA 4 001 40 00 M 0 01M 2 10 40 01 400 0 M 0 1M 2 5 400 1 4000 M 1M 4 5 1000V 0...

Page 19: ...for positive Minus sign for negative Measurement Rate 2 times per second nominal Low Battery Indication is displayed if battery voltage drops below operating voltage Battery Six 6 1 5V AA batteries F...

Reviews: