10 

General Specifications 

Diode Test

 

Bias voltage: 2.8VDC typical 

Continuity Check

 

Audible signal will sound if the resistance is less than 100

Ω

 

Input Impedance

 1M

 (VDC & VAC) 

AC Bandwidth 

50 / 60Hz 

Display 

2000 count (0 to 1999) LCD 

Overrange indication

 

“1___” is displayed 

Polarity

 

Automatic (no indication for positive); Minus (-) sign for 
negative 

Measurement Rate 

2 times per second, nominal 

Low Battery Indication 

 ”

 is displayed  

Battery

 

One (1) 9V battery (NEDA 1604)

 

Fuses

  

mA range; 200mA/660V ceramic fast blow 
A range; 10A/250V ceramic fast blow 

Operating Temperature

 

0ºC to 50ºC (32ºF to 122ºF) 

Storage Temperature

  

-20

o

C to 60

o

C (-4

o

F to 140

o

F) 

Operating Humidity 

Max 70% up to 31ºC (87ºF) decreasing linearly to 50% at 50ºC   
(122ºF) 

Storage Humidity 

< 80% RH 

Operating Altitude 

2000 meters. (7000 ft) maximum. 

Weight

  

255g (9.17 oz)  

Size

  

132 x 66 x 38mm (5.2” x 2.6” x 1.5”) 

Approvals 

CE 

Safety

  

This meter is intended for indoor use and protected, against 
the users, by double insulation per EN61010-1 and IEC61010-
1 2nd Edition (2001) to CAT II 1000V & CAT III 600V; Pollution 
Degree 2.  

PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY 

OVERVOLTAGE CATEGORY I 

Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in 
which measures are taken to limit the transient overvoltages to an appropriate low level. 
Note – Examples include protected electronic circuits. 

OVERVOLTAGE CATEGORY II 

Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be 
supplied from the fixed installation. 
Note – Examples include household, office, and laboratory appliances. 

OVERVOLTAGE CATEGORY III 

Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. 
Note – Examples include switches in the fixed installation and some equipment for 
industrial use with permanent connection to the fixed installation. 

OVERVOLTAGE CATEGORY IV 

Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. 
Note – Examples include electricity meters and primary over-current protection equipment 

 
 

©Copyright 

©

 2008 Extech Instruments Corporation (a Flir company) 

All rights reserved including the right of reproduction in whole or in part in any form. 

Summary of Contents for Extech Instruments MN15

Page 1: ...User s Guide Mini Digital Multimeter Model MN15...

Page 2: ...r manual before operating the meter Always remove the test leads before replacing the battery or fuses Inspect the condition of the test leads and the meter itself for any damage before operating the...

Page 3: ...ntinuity 250Vrms for 15 sec max Controls and Jacks 1 LCD display 2 Rotary function switch 3 10 ampere test lead jack 4 Test lead jack for voltage milliamp resistance continuity diode and temperature f...

Page 4: ...eing switched ON or OFF Large voltage surges may occur that can damage the meter 1 Set the function switch to the 600 VAC position 2 Insert the black test lead banana plug into the negative COM jack I...

Page 5: ...the 200mA position and insert the red test lead banana plug into the mA jack 3 For current measurements up to 10A set the function switch to the 10A position and insert the red test lead banana plug...

Page 6: ...e potential 1 Set the function switch to the position 2 Insert the black lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive jack 3 Touch the test probe...

Page 7: ...e positive V jack 3 Touch the black test probe tip to the negative side of the battery Touch the red test probe tip to the positive side of the battery 4 Read the voltage in the display TEMPERATURE ME...

Page 8: ...BE STORED FOR A LONG PERIOD OF TIME the battery should be removed to prevent damage to the unit BATTERY FUSE INSTALLATION and LOW BATTERY INDICATION WARNING To avoid electric shock disconnect the tes...

Page 9: ...ifications Function Range Resolution Accuracy 200mV 0 1mV 2000mV 1mV 20V 0 01V 0 5 reading 2 digits 200V 0 1V DC Voltage V DC 600V 1V 0 8 reading 2 digits 200V 0 1V AC Voltage V AC 600V 1V 1 2 reading...

Page 10: ...er EN61010 1 and IEC61010 1 2nd Edition 2001 to CAT II 1000V CAT III 600V Pollution Degree 2 PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equipment of OVERVOLTAGE CATEGORY I is...