other optical and geometrical disturbances
3
. For more information, see section 30
, page 118.
An NUC is performed automatically, for example at start-up, when changing a measure-
ment range, or when the environment temperature changes.
You can also perform an NUC manually. This is useful when you have to perform a critical
measurement with as little image disturbance as possible. You may, for example, want to
perform a manual calibration just before you start recording a video sequence.
11.8.2
Performing an NUC manually
1. To perform a manual NUC, push and hold down the image archive button
for
more than 2 seconds.
Note
You can also assign the function
Calibrate
to the one of the programmable but-
tons. For more information, see section 21.15
Assigning functions to the programmable
11.9
Hiding all overlay
11.9.1
General
The camera overlay consists of overlay graphics and image overlay information. The
overlay graphics include items such as measurement tool symbols, result tables, and
status icons. The image overlay information, which you activate on the
Settings
menu,
provides additional information such as the date, time, and items related to temperature
measurements. For more information, see section 9.7.6
You can hide all camera overlay by touching the soft button
.
Note
You can also assign the function
Hide image overlay graphics
to one of the pro-
grammable buttons. For more information, see section 21.15
#T810411; r. AC/70277/70566; en-US
33
3. Definition from the European standard EN 16714-3:2016, Non-destructive Testing—Thermographic Testing—
Part 3: Terms and Definitions.
Summary of Contents for GF7 Series
Page 1: ...User s manual FLIR GF7x series...
Page 93: ...Handling the camera 21 21 8 2 Figure T810411 r AC 70277 70566 en US 85...
Page 119: ...Mechanical drawings 28 See next page T810411 r AC 70277 70566 en US 111...
Page 124: ...CE Declaration of conformity 29 See next page T810411 r AC 70277 70566 en US 116...
Page 125: ......