FLIR
IM75
USER
GUIDE
Document
Identifier:
IM75
‐
en
‐
US_AA
24
Function
Resistance
Range Resolution
Accuracy
of
Reading
Insulation
Resistance
4.000M
Ω
40.00
M
Ω
0.001M
Ω
0.01M
Ω
±
(1.5%
+
5
digits)
400.0M
Ω
4000
M
Ω
0.1M
Ω
1M
Ω
±
(3.0%
+
5
digits)
4.1G
Ω‐
20.0G
Ω
0.1G
Ω
±
(10.0%
+
3
digits)
Test
voltages
Min.
Resistance
Max.
Resistance
50V
50k
Ω
50.0M
Ω
100V
100k
Ω
100.0M
Ω
250V
250k
Ω
250.0M
Ω
500V
500k
Ω
500.0M
Ω
100V
1M
Ω
20G
Ω
Notes:
1.
Test
Current
=
1
mA
2.
Test
Voltage
Accuracy
=
0%,
+20%
Short
Circuit
Test
Current
=
1mA
(nominal)
3.
Auto
Discharge
Function:
Discharge
time
<1
second
for
Cap.
<1µF
4.
Max.
Capacitive
Load:
Operable
with
up
to
1µF
load.
5.
Live
Circuit
Detection:
If
>2V
AC/DC
detected
test
inhibited
7.3
Input
Specifications
Function
Maximum
Input
AC
Voltage,
DC
Voltage
1000V
DC/AC
Insulation
Test
No
LIVE
Voltage
IN
‐
Protected