24
intelliRock® III– Operating Instructions Guide
Note: for maturity applications, it is not necessary to upload the calibration curve for the job to
each logger at logger startup. Quite often, the calibration curve is created using the same mix
being poured in the placement. Once a calibration curve for a mix is ready, follow the instruc-
tions found in a later section of this guide to use that calibration curve with loggers on a spe-
cific job. If a calibration curve is available for use with a logger, press F2 to associate it with the
currently connected logger.
Resetting a Logger
The first 60 minutes after starting a logger is an interval during which a user can reset the log-
ger. After this interval is past, the logger becomes locked and can no longer be restarted.
To reset a logger during this period, from the screen shown in Figure 4.8, select the “Logger”
(F1) menu option. The screen shown in Figure 4.9 will appear.
Figure 4.9: Restarting a logger.
From the screen in Figure 4.9, press the “Shift” key and the “Reset” (F1) key simultaneously. The
screen will temporarily display “Updating” before returning to the first logger setup screen.
Reviewing Logger Data
To review a logger’s data using the intelliRock reader, connect the logger wires to the reader.
When a logger is successfully connected to the reader, the screen will temporarily display the
“Good Link” message and the serial number of the connected logger as shown in Figure 4.10.
Figure 4.10: The reader screen showing a successful logger connection.