6.3.2.1
Installing an infrared lens
Note
• The detector is a very sensitive sensor. It must not be directed toward strong visible
light, e.g., sunlight.
• Do not touch the lens surface when you install the lens. If this happens, clean the lens
according to the instructions in section 12.1.2
• Do not touch the filter surface when you install the lens. If this happens, clean the filter
according to the instructions in section 12.1.2
6.3.2.1.1
Procedure—bayonet mount
Follow the procedure below to install an infrared lens with a bayonet mount:
1. If present, remove the installed lens or the protection in front of the detector/filter
wheel.
2. Align the red index mark on the lens with the red index mark on the bayonet ring.
3. Carefully push the infrared lens into the bayonet ring.
4. Rotate the infrared lens 30° clockwise (looking at the front of the lens).
5. The camera automatically identifies the lens and selects the measurement configura-
tion corresponding to the lens, if loaded into the camera memory. For more informa-
tion, see section 6.4.4
, page 23.
6.3.2.1.2
Procedure—M80 mount
Follow the procedure below to install an infrared lens with an M80 mount:
1. If present, remove the installed lens or the protection in front of the detector/filter
wheel.
2. Carefully push the infrared lens into position.
3. Rotate the infrared lens clockwise (looking at the front of the lens) until it stops.
4. You must manually select the measurement configuration using the FLIR ResearchIR
Max software. For more information, see section 6.4.4
page 23.
6.3.2.2
Removing an infrared lens
Note
• The detector is a very sensitive sensor. It must not be directed toward strong light, e.
g., sunlight.
• Do not touch the lens surface when you install the lens. If this happens, clean the lens
according to the instructions in section 12.1.2
• Lenses can be heavy, so take care not to be surprised by their weight. Some lenses
weight several hundred grams.
• When you have removed the infrared lens, put the lens caps on the lens to protect it
from dust and fingerprints.
6.3.2.2.1
Procedure—bayonet mount
Follow the procedure below to remove an infrared lens with a bayonet mount:
1. Pull forward the release button for the infrared lens (4 in Figure 5.1
2. Rotate the infrared lens 30° counterclockwise (looking at the front of the lens).
3. Carefully pull out the infrared lens from the bayonet mount.
4. Install the protective cap or a new optic on the camera, to avoid visible light striking
the detector.
6.3.2.2.2
Procedure—M80 mount
Follow the procedure below to remove an infrared lens with an M80 mount:
1. Rotate the infrared lens counterclockwise (looking at the front of the lens).
2. Carefully pull out the infrared lens.
#T810206; r. AA/43064/43088; en-US
19
Summary of Contents for X6530sc Series
Page 1: ...User s manual FLIR X6530sc series...
Page 2: ......
Page 3: ...User s manual FLIR X6530sc series T810206 r AA 43064 43088 en US iii...
Page 4: ......
Page 65: ...Mechanical drawings 10 See next page T810206 r AA 43064 43088 en US 57...
Page 117: ......