Synchronization parameters are set through the ResearchIR Max user interface:
1. Activate/deactivate external synchronization. Select the active edge and input
impedance.
2. Based on the camera configuration (e.g., the window size, integration time, or inte-
gration mode), the maximum allowable Sync In frequency is displayed.
3. The actual Sync In signal frequency is measured by the camera and displayed here.
If the Sync In frequency is higher than the maximum allowable frame rate, a warning
message is displayed. In this case, the input signal is under-sampled.
4. The jitter on the Sync In signal, which is typically one pixel clock, is displayed here.
5. The integration time length is displayed here. The integration time is defined in the
measurement range panel.
6. A delay between the Sync In signal and the start of integration time can be defined
here.
7. Several camera signals can be routed to the Sync Out connector. The polarity of
these signals is also defined here.
6.4.8
Advanced camera controls
6.4.8.1
General
This section describes the
Advanced Camera Controls
.
#T810206; r. AA/43064/43088; en-US
26
Summary of Contents for X6530sc Series
Page 1: ...User s manual FLIR X6530sc series...
Page 2: ......
Page 3: ...User s manual FLIR X6530sc series T810206 r AA 43064 43088 en US iii...
Page 4: ......
Page 65: ...Mechanical drawings 10 See next page T810206 r AA 43064 43088 en US 57...
Page 117: ......