Thermographic measurement techniques
5. Measure the apparent temperature of the aluminum foil and write it down. The foil is
considered a perfect reflector, so its apparent temperature equals the reflected appa-
rent temperature from the surroundings.
Figure 16.5
Measuring the apparent temperature of the aluminum foil.
16.2.1.2
Step 2: Determining the emissivity
Follow this procedure:
1. Select a place to put the sample.
2. Determine and set reflected apparent temperature according to the previous
procedure.
3. Put a piece of electrical tape with known high emissivity on the sample.
4. Heat the sample at least 20 K above room temperature. Heating must be reasonably
even.
5. Focus and auto-adjust the camera, and freeze the image.
6. Adjust
Level
and
Span
for best image brightness and contrast.
7. Set emissivity to that of the tape (usually 0.97).
8. Measure the temperature of the tape using one of the following measurement
functions:
•
Isotherm
(helps you to determine both the temperature and how evenly you have
heated the sample)
•
Spot
(simpler)
•
Box Avg
(good for surfaces with varying emissivity).
9. Write down the temperature.
10. Move your measurement function to the sample surface.
11. Change the emissivity setting until you read the same temperature as your previous
measurement.
12. Write down the emissivity.
Note
• Avoid forced convection
• Look for a thermally stable surrounding that will not generate spot reflections
• Use high quality tape that you know is not transparent, and has a high emissivity you
are certain of
• This method assumes that the temperature of your tape and the sample surface are
the same. If they are not, your emissivity measurement will be wrong.
#T810206; r. AA/43064/43088; en-US
78
Summary of Contents for X6530sc Series
Page 1: ...User s manual FLIR X6530sc series...
Page 2: ......
Page 3: ...User s manual FLIR X6530sc series T810206 r AA 43064 43088 en US iii...
Page 4: ......
Page 65: ...Mechanical drawings 10 See next page T810206 r AA 43064 43088 en US 57...
Page 117: ......