User Manual X8400 sc
38
LED Description
LED STATUS
Description
OFF
No signal detected. Check connection
and signal levels
GREEN
Signal is detected and signal voltage is
correct, but signal is continuous
ORANGE
Signal is detected but signal voltage is
incorrect
Blinking Green
Signal is detected.
LED is blinking at signal frequency.
Signal voltage is correct
Blinking Orange
Signal is detected.
LED is blinking at signal frequency.
Signal voltage is incorrect
6.7.2 Sync Out
General
The Sync OUT signal is synchronous with the Sync IN or with the frame rate (if the Sync In is
not selected). It can be used to synchronize other events with the camera. It is a TTL signal.
The Sync OUT setup is described in §4.4.7
Characteristics
Name
Value
Amplitude (V)
TTL signal 0/+5 V
Max frequency (Hz)
Maximum Frame Rate of the camera for a
given detector configuration
Impedance
High impedance
Minimum Pulse width (ns)
300 ns
Protection
Voltage peaks (500
V / <1
ns)
Overvoltage (15 V)
Reversed polarity
Connector
Coaxial BNC Jack
LED Description
LED Status
Description
OFF
No signal.
Blinking GREEN
Signal is ready to use.
LED is blinking at signal frequency.
GREEN
Signal is not usable because it
saturates the camera input.
Signal voltage is 5V continuously
.