background image

  

 

 

 

 

Installation   9 

 

2.4     Meters fitted with integral Instruments  

If your flowmeter is fitted with an integral instrument such as a 

totaliser, rate totaliser or batch controller then the pulse output from the meter has been factory wired to the flow 
input of the readout instrument.

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
As a default the reed output is pre-wired and DIP switches set for a integral totaliser or rate/totaliser allowing self 
powered operation of the instrument displays. 
 
Also by default the open collector output from the Hall Sensor is pre-wired and DIP switches set for a integral batch 
controller allowing high speed, solid state operation of the model EB batch controller.  
 
These defaults may vary at the customer request or for specific applications such as dual flow input or high or low 
flow so if unsure remove the instrument bezel to check the wiring. 
 
The output(s) and function(s) available from a meter fitted with an integral instrument depends on the model of the 
instrument fitted and may include meter pulse repeater, prescaled pulse output, 4-20mA flow output, flowrate 
alarms or single/dual stage batch control logic 

(preset controller).

  

 
Refer to the option in the meter model number and relevant instrument manual. Unless programming details were 
provided at time of order the instrument program will contain factory default parameters. Integral instruments will 
however be programmed with the relevant calibration factor 

(K factor or scale factor) 

for the meter.  

 
Factory default settings can be found in the instrument instruction manual and it should be noted all output(s) are 
turned OFF and if required need to be turned ON then programmed to suit the application requirements. 

 
 

2.4.1    Meter Calibration Factor 

(K or scale Factor)   

Each flowmeter is individually calibrated and supplied with a 

calibration certificate showing the number of pulses per unit volume 

(eg pulses per litre or pulses per USgallon).

 

Nominal figures are shown in the specification section of this manual.

 

 
Meters fitted with Integral Instruments will have the relevant calibration factor entered into the program of the 
instrument. Please refer to relevant instrument manual for programming details. 

10      Commissioning & Maintenance  

 
3.0     Commissioning   

Once the meter has been mechanically and electrically installed in accordance with this 

and any other relevant instrument manual(s) the meter is ready for commissioning.

 

 

The meter must NOT be run until the pipework is flushed of foreign matter,

 

more often than not  foreign matter is 

present after pipework fabrication or modification, weld slag, grinding dust, sealing tape & compound &/or surface 
rust are most common offenders.

  

 

Flushing can be undertaken by utilizing a by-pass or removing the meter from the pipework. If neither is practical 
then the meter rotors must be removed prior to flushing 

(refer to Maintenance section of this manual for 

disassembly).

 

  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

After flushing or following long periods of shutdown the meter must be purged of air/vapour. This can be achieved 
by allowing the liquid to flow through the meter at a slow rate until all air/vapour is displaced. Never run the meter 
above its maximum flow or exceed 100kpa (1 bar, 15psi) pressure drop across the meter. Now the meter is ready 
for its operation to be confirmed by ensuring correct indication or operation at the receiving instrument(s). Refer if 
necessary to fault finding section of this manual. 

 
 

4.0     Maintenance   

Adhering to the installation instructions in this manual should ensure your meter provides the 

required operational performance. These are mechanical meters and a periodic maintenance and inspection 
regime will maximize the operational availability of the meter.

 

 
The frequency of maintenance depends on the application factors including liquid lubricity and abrasiveness and 
operational factors such as flowrate and temperature. 
 

BEFORE

 

undertaking meter maintenance ensure the following :

  

Associated alarm(s) or control output(s) are isolated so not to affect the process. 

Voltage supply is isolated from the meter. 

Liquid supply to the meter is closed off. 

The meter is depressurized and liquid drained from the meter.  

strainer 

flowmeter 

isolation valve 

isolation valve 

by-pass valve 

Open downstream valve last  

with LCD 

register 

Summary of Contents for OM025

Page 1: ...w flomec com au USA 916 Belcher Drive Pelham AL 35124 USA Tel 1 205 378 1050 Fax 1 205 685 3001 Email customerservice trimecus com Website www trimecus com Distributed by Each meter has been calibrated on mineral oil and will contain a small amount of oil residue The oil used is Castrol Diesel Calibration Fluid 4113 product code 055830 Medium capacity positive displacement flowmeters OVAL GEAR SEE...

Page 2: ...otalisers rate totalisers or batch controllers These instruments also have monitoring and control output options including 4 20mA scaled pulse flowrate alarms and batch control logic preset metering If your flowmeter is fitted or supplied with an instrument please also refer to the relevant instrument instruction manual These flowmeters can be installed within hazardous areas by using the reed swi...

Page 3: ...en collector phased outputs 4 dc 2 stage batch controller OM040 OM050 accum reset totals pulse output IECEX ATEX approved IECEX ATEX approved IECEX ATEX approved Specifications Model prefix OM015 OM025 OM040 OM050 Nominal size inches 15mm 1 2 25mm 1 40mm 11 2 50mm 2 Flowrange litres min 1 40 10 150 15 250 30 450 Flowrange USGM 0 26 10 6 2 6 40 4 66 8 120 Accuracy 3cp Repeatability Temperature rang...

Page 4: ...nd air entrapment which can result in erroneous readings or damage the meter on start up Process or safety critical meters should be installed in a by pass section of pipe with isolation valves to enable the meter to be isolated and serviced as required A by pass installation also allows purging of the system during commissioning see Commissioning The meter must be appropriately rated and is typic...

Page 5: ...stor typically greater than 10K ohms in most instruments which ties the open collector to the available dc voltage level when the Hall sensor is not energized When energized the open collector output is pulled to ground through the emitter 0V 2 3 2 Reed Switch Pulse Output The reed switch output is a two wire normally open SPST voltage free contact ideal for installations without power or for use ...

Page 6: ...ill have the relevant calibration factor entered into the program of the instrument Please refer to relevant instrument manual for programming details 10 Commissioning Maintenance 3 0 Commissioning Once the meter has been mechanically and electrically installed in accordance with this and any other relevant instrument manual s the meter is ready for commissioning The meter must NOT be run until th...

Page 7: ...embly with rotor shafts Part No aluminum 1401096 1401094 1401092 stainless steel 1401097 1401093 2 Rotor assembly set aluminum rotors bearings 1424071 1424111 1424099 stainless steel rotors bearings 1424092 1424125 1524011 3 Body O ring size BS153 size BS245 size BS253 viton standard 13031531 13032451 13032531 teflon 13031533 13032453 13032533 4 Meter cap aluminum with M20 conduit entry 1302047 13...

Page 8: ...tep 1 Check application installation and set up Refer to Mechanical Installation section for installation and application factors that may effect the meter operation including pulsation and air entrainment or incorrect meter selection including incorrect flow rate temperature and pressure or materials compatibility Refer to Electrical Installation for correct wiring Step 2 Check for blockages The ...

Page 9: ...rminations wires for continuity 1 Rotors fouled 1 Check that rounded teeth are towards base of chamber 2 Check for obstruction due to foreign particles 3 Clean repair or replace rotors 2 1 See instructions for reassembly of meter with particular emphisis on positioning of rotors magnets 3 1 Check terminal connections solder joints 2 Ensure dc voltage is available at Vdc 0v and receiving instrument...

Reviews: