  Trouble shooting     15 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 

            

Notes 

5.1   TROUBLE SHOOTING

Symptom

Solution

1.

1. Ground shield of signal cable
2. Re-route cable from high electrical energy sources

2.

1. Remove source of air or gas entrapment

2. Install an upstream air eliminator 

3.

1. Increase back pressure on pump

2. Install a fast response one way check valve

3. Install a surge arrestor between pump & meter

4. Re-calibrate meter in situ to compensate for pulsations

5. Change pump style to smooth delivery type pump

1.

Damaged or

1. Inspect, repair, clean or replace rotors

worn rotors

2.

1. Inspect measuring chamber for damage - repair

2. Check concentricity of rotor shafts within chamber

3.

1. Ground shield of signal cable
2. Re-route cable from high electrical energy sources

3. Check all electrical terminations & wires for continuity. 

1.

Rotors fouled

1. Check that rounded teeth are towards base of chamber

2. Check for obstruction due to foreign particles

3. Clean, repair or replace rotors

2.

1. See instructions for reassembly of meter with

particular emphisis on positioning of rotors & magnets

3.

1. Check terminal connections & solder joints

2. Ensure dc voltage is available at Vdc & 0v and

receiving instrument is fitted with a pull up resistor

3. Replace output board

1.

1. Check DIP switch settings & program data

2. Check terminal connections & electrical continuity

3. Repair / replace receiving instrument

Possible cause

Output signal 

interference

Pulsating flow from 

reciprocating style 

pump

Damaged or worn 

measuring 

chamber

No output from 

output board

Faulty receiving 

instrument

Meter incorrectly 

reassembled

Meter 

readings 

are high

Output signal 

interference

Meter 

readings 

are low

Entrained air or 

gas

No output 

from meter

Not reading 

on readout 

instrument

Summary of Contents for OM025

Page 1: ...w flomec com au USA 916 Belcher Drive Pelham AL 35124 USA Tel 1 205 378 1050 Fax 1 205 685 3001 Email customerservice trimecus com Website www trimecus com Distributed by Each meter has been calibrated on mineral oil and will contain a small amount of oil residue The oil used is Castrol Diesel Calibration Fluid 4113 product code 055830 Medium capacity positive displacement flowmeters OVAL GEAR SEE...

Page 2: ...otalisers rate totalisers or batch controllers These instruments also have monitoring and control output options including 4 20mA scaled pulse flowrate alarms and batch control logic preset metering If your flowmeter is fitted or supplied with an instrument please also refer to the relevant instrument instruction manual These flowmeters can be installed within hazardous areas by using the reed swi...

Page 3: ...en collector phased outputs 4 dc 2 stage batch controller OM040 OM050 accum reset totals pulse output IECEX ATEX approved IECEX ATEX approved IECEX ATEX approved Specifications Model prefix OM015 OM025 OM040 OM050 Nominal size inches 15mm 1 2 25mm 1 40mm 11 2 50mm 2 Flowrange litres min 1 40 10 150 15 250 30 450 Flowrange USGM 0 26 10 6 2 6 40 4 66 8 120 Accuracy 3cp Repeatability Temperature rang...

Page 4: ...nd air entrapment which can result in erroneous readings or damage the meter on start up Process or safety critical meters should be installed in a by pass section of pipe with isolation valves to enable the meter to be isolated and serviced as required A by pass installation also allows purging of the system during commissioning see Commissioning The meter must be appropriately rated and is typic...

Page 5: ...stor typically greater than 10K ohms in most instruments which ties the open collector to the available dc voltage level when the Hall sensor is not energized When energized the open collector output is pulled to ground through the emitter 0V 2 3 2 Reed Switch Pulse Output The reed switch output is a two wire normally open SPST voltage free contact ideal for installations without power or for use ...

Page 6: ...ill have the relevant calibration factor entered into the program of the instrument Please refer to relevant instrument manual for programming details 10 Commissioning Maintenance 3 0 Commissioning Once the meter has been mechanically and electrically installed in accordance with this and any other relevant instrument manual s the meter is ready for commissioning The meter must NOT be run until th...

Page 7: ...embly with rotor shafts Part No aluminum 1401096 1401094 1401092 stainless steel 1401097 1401093 2 Rotor assembly set aluminum rotors bearings 1424071 1424111 1424099 stainless steel rotors bearings 1424092 1424125 1524011 3 Body O ring size BS153 size BS245 size BS253 viton standard 13031531 13032451 13032531 teflon 13031533 13032453 13032533 4 Meter cap aluminum with M20 conduit entry 1302047 13...

Page 8: ...tep 1 Check application installation and set up Refer to Mechanical Installation section for installation and application factors that may effect the meter operation including pulsation and air entrainment or incorrect meter selection including incorrect flow rate temperature and pressure or materials compatibility Refer to Electrical Installation for correct wiring Step 2 Check for blockages The ...

Page 9: ...rminations wires for continuity 1 Rotors fouled 1 Check that rounded teeth are towards base of chamber 2 Check for obstruction due to foreign particles 3 Clean repair or replace rotors 2 1 See instructions for reassembly of meter with particular emphisis on positioning of rotors magnets 3 1 Check terminal connections solder joints 2 Ensure dc voltage is available at Vdc 0v and receiving instrument...