27
OPERATION (continued)
USER CONFIGURATION MODE (Field Calibration Mode - continued)
Percent Adjust Method (Correction Factor)
Percent adjust method is a field calibration method by which the user will enter a
percentage amount to adjust K-factor values in the factory programmed K-factor table.
Once the percentage is input by the user, the software will apply this percentage
adjustment to all calculated and displayed values. (See Figure 33-1)
This will not over write the existing factory
calibration table, instead, it applies a
calculation to the existing values for use by
the computer.
This method is good for users who simply
want to adjust their meter by a percentage
amount.
Figure 33-1
The top row left most character will be in
focus to indicate the user can change the
value. The left most character is the sign of
the value. The decimal point is not
adjustable. The range for the values are: -
99.999% to +99.999%. The software will
display the user applied percent
adjustment on the top row.
Figure 33-2
Momentarily press
Rate
to toggle between (dash “-“) and (underscore “_”). A dash
indicates a negative value. An underscore indicates a positive value. Press
Total
to
edit the adjustment value.
When all digits are entered correctly, exit by holding the
Total
and pressing the
Rate
button. The software will store the adjusted percentage correction, then display the
“Save” screen. (See Figure 13)
NOTE:
Failure to save your entries on the “save” screen will cause the entries to be
discarded.
K-factor Entry Method
K-factor entry method (see Figure 34-1) is a field calibration method by which the user
can input either a single point or multipoint calibration table.
From 1 to 5 unique K-factor and frequency
values can be entered and will be stored in
the custom K-factor table in the computer.
This will not over write the existing factory
calibration table.
Figure 34-1