Usually the necessary data is provided with the flowme-
ter. If not, your flowmeter must be calibrated for use.
This can be done by selecting Digital Linear in hardware
and setting your K factor to 1 in variables. By making
various system runs with calibrated measurements, the
actual flow verses the frequency input can be determined
and K factors derived.
Below is some sample data that compensates for some low
end inaccuracies. This sample is an illustrative example
applicable to a gas application using English units.
Sample Data:
Input
Actual ft
3
/sec
Frequency
Actual flow
K factor
0
0
1.0000
10
8
1.2500
20
18
1.1111
30
29.5
1.0170
100
100
1.0000
1000
1000
1.0000
The data would be entered as follows
Display shows
Press
Enter Point
Freq01?
00
Freq01? 0.0000
K Factor01?
1
Enter Point
Freq02?
10
Freq02? 10.0000
K Factor02?
1.25
Enter Point
Freq03?
20
4-3