background image

 

 

 

 

 

Troubleshooting APEX 8000  Positioners

 

Failure 

Probable Cause 

Corrective Action 

Valve won’t 

stroke, no 

excessive air 
is exhausting 

from 

positioner 

1.  Tubing to wrong ports 

2.  Cam action reversed 
3.  Feedback lever arm is stuck 
4.  low air supply 
5.  Relay tube stuck 

 

6.  Balance adjust screw not adjusted correctly 
7.  I/P module filter plugged 

8.  I/P module failure 
9.  I/P mounting bolts loose 

10.I/P pressure signal blocked 

1.   Re-tube to correct ports (see “installation section). 
2.   Refer to “installation” section and reverse cam 
3.   Work with lever arm until it turns freely 
4.   Increase air supply to recommended value 
5.   Disassemble relay assembly and work relay tube free. Lightly   lubricate 

if necessary 

6.   Adjust balance pressure with adjusting screw 
7.   Remove I/P module and replace filter 
8.   Replace I/P module 
9.   Tighten mounting bolts 
10. Remove I/P module and clear passageway; replace O-ring if necessary 

 

Valve  won’t 

stroke, 

excessive air 

exhausting 

from 

positioner 

1.  A diaphragm in relay assembly burst 
2.  One of the poppets is stuck 
3.  Internal control valve problem 
4.  Damaged relay O-rings on relay tube 
5.  Blocked passageways in relay 

1.  Replace relay assembly or replace diaphragms 
2.  Remove relay assembly/poppet cover; free stuck poppet 
3.  Refer to instructions or check for actuator tubing leaks 
4.  Disassemble relay and replace O-rings 
5.  Disassemble relay and check small holes under diaphragms; clean if 

clogged 

Actuator goes 

to full signal 

position 

regardless of 

signal 

1.  Broken feedback spring 
2.  Linkage is disconnected or stuck 
 
3.  Orifice is clogged by water, oil or dust in air 

supply 

4.  Bent flapper, damaged nozzle 
5.  I/P module failure 
6.  Clogged orifice screen 

1.  Replace feedback spring 
2.  Check and tighten bolts/nuts in linkage. Make sure linkage dones not 

hang up; grease ping that rides in follower arm slot 

 
3.  Remove orifice screw and carefully clean orifice hole 
4.  Straighten flapper or replace damaged parts 
5.  Replace I/P module 
6.  Remove relay and clean ore replace orifice screen 

Calibration 1. 

Loose 

positioner mounting 

 

2. Loose linkage 
3.  Loose zero adjustment locking knobs 
4.  Wear of arms or pins 
5.  I/P mounting bolts loose 
6.  Stroke has change in valve 

1.  Remove cover and check three screws holding positioner to bracket; 

check two bolts holding bracket to yoke 

2.  Tighten nuts and bolts on linkage and stem clamp 
3.  Tighten zero adjustment locking knob; re-calibrate if necessary 
4.  Replace worm arms, pins; grease appropriately 
5.  Tighten I/P mounting bolts 
6.  Refer to valve maintenance instructions 

Excessive air 

consumption 

(other than 

normal 

exhaust) 

1.  Air leakage from manifold rings between 

relay and base 

2.  Air leakage from tubing 
3.  Leaky cylinder piston O-rings 
4.  Air leakage from relay 

1.  Tighten screws holding relay assembly together and/or replace O-rings 
 
2.  Tighten or replace tubing fittings 
3.  Replace O-rings in cylinder 

4.  Disassemble relay and check and replace dynamic O-rings next to tube 

if necessary 

Actuator 

strokes very 

slowly in one 

direction only 

1.  Connection between signal capsule and 

flapper misadjusted 

 
 

2.  Tubing to cylinder is restricted 
3.  Balance pressure low 

1.  Adjust gain according to Figure 6 or until actuator strokes approximately 

equal speed in both directions. Verify alignment of upper and lower gain 
plates. Make sure spacer nut is tightened. 

2.  Inspect tubing/fittings for restriction and replace if necessary 
3.  Adjust balance pressure according to page 5 

Erratic 

operation 

1.  Dirt buildup on relay poppets or seats 
2.  Dirt buildup on relay tube 
3.  Clogged ports / passageways in relay 
4.  Faulty I/P module 
5.  Clogged orifice screw 
6.  Mechanical binding in linkage or internal 

galling in valve 

7.  Clogged orifice screen 

1.  Disassemble; clean poppets and seats; add air filter or change filter 
2.  Disassemble; clean relay and lightly lubricate; replace O-rings if 

necessary; add air filter or change filter 

3.  Disassemble, inspect and clean all ports and passageways 
4.  Replace the I/P module 
5.  Remove orifice screw and carefully clean orifice 
6.  Tighten linkage or refer to valve maintenance instructions 
 
7.  Remove relay and clean or replace orifice screen 

Excessive 

overshoot 

1.  Restriced air flow to positioner 
2.  Balance pressure not set correctly 
3.  Gain is set too high 

1.  Adjust air supply as needed 
2.  Adjust balance pressure according to page 5 
3.  Lower gain mechanism until overshoot is minimized 

 

 

Summary of Contents for APEX 8000 Series

Page 1: ...lugging one of the output ports NOTE The APEX 8000 high performance positioner must use the I P NT 3000 Transducer The I P 2000 Transducer is not acceptable for use with the APEX 8000 Series Positioner The APEX 8000 positioner can handle supply pressures up to 150 psi thus a supply regulator is usually not required However a five micron air filter is required for pneumatic positioners and a coales...

Page 2: ... capsule moves up it will close the exhaust seat of the upper pilot poppet and open the supply seat which applies increased air pressure to the bottom cylinder port At the same time the pilot valve capsule will open the exhaust seat for the lower pilot poppet thus decreasing pressure to the top cylinder port This difference in pressure will drive the piston outward rotate the pinion and stretch th...

Page 3: ...ositioner onto the bracket Make sure positioner shaft and coupler are engaged and centered Finger tighten positioner bolts 5 Proceed to Connecting APEX 8000 Positioner Ports Connecting APEX 8000 Positioner Ports 1 For double acting actuators connect positioner ports 1 and 2 to the actuator Port 1 is always connected to the actuator port used to drive the actuator away from its start or fail positi...

Page 4: ... if necessary 6 Connect supply pressure to port marked Supply stroke actuator valve two or three times to align position with actuator With a 50 percent input signal tighten all mounting bolts Stroke actuator to verify proper alignment Position Calibration Introduction APEX 8000 positioners are calibrated at the factory however due to shipping and handling it may be necessary to check the calibrat...

Page 5: ... Cam return spring See Figure 4 Positioner Balance Adjustment CAUTION Balance pressure is 75 percent of the supply pressure and is the average pressure of output ports 1 and 2 Balance is preset at factory However if this adjustment becomes necessary carefully make this adjustment slowly allowing the positioner to stabilize Check after a short while to make sure balance pressure is correct Balance ...

Page 6: ... travel stops Allow actuator pressure to stabilize 7 Observe the pressure gauges If reading is not correct turn balance adjustment screw about 1 8 turn at a time and wait about 20 30 seconds for pressure to stabilize counterclockwise to increase pressure Continue until output pressure level of the higher pressure gauge is approximately 80 percent of supply 8 Replace rubber cap over balance adjustm...

Page 7: ...k actuator responsiveness by providing a step signal to positioner When gain is set as desired check valve zero and span calibration and re calibrate if needed 7 Return the valve to service Calibrating I P Module Zero and Span Settings NOTE Although calibration can be accomplished using the output pressure gauge on the I P module its accuracy is 3 percent The standard gauge should be removed only ...

Page 8: ...er adjustments are obtained Adjusting the Minimum Pressure Cutoff Feature The APEX 8000 positioner with I P Transducer has a Minimum Pressure Cutoff MPC feature which allows the user to set the positioner When the input signal falls below a user adjustable current the pressure output falls rapidly to approxi mately 1 7 psi causing the valve to move to the failure position This feature is generally...

Page 9: ...o ensure a clean air supply Check and maintain filter regularly 2 Make sure all arms and levers move freely 3 Check for loose parts 4 Be sure there are no leaks in the air supply tubing fittings or connections 5 Refer to the troubleshooting chart on page 9 in case of problems Pilot Relay Disassembly and Reassembly The pilot relay is available as a complete unit and can be easily replaced See Steps...

Page 10: ...pper relay body 9 being careful not to damage seating surface or O rings 14 Reinstall poppets 28 poppet springs 29 Seat Spring 39 O rings 30 and poppet covers 27 before installing retaining rings 31 15 If signal diaphragm assembly 16 is damaged proceed as follows With relay halves still apart remove four screws 32 holding signal diaphragm assembly 16 to the pilot relay assembly Remove locking scre...

Page 11: ...t locking knobs 4 Wear of arms or pins 5 I P mounting bolts loose 6 Stroke has change in valve 1 Remove cover and check three screws holding positioner to bracket check two bolts holding bracket to yoke 2 Tighten nuts and bolts on linkage and stem clamp 3 Tighten zero adjustment locking knob re calibrate if necessary 4 Replace worm arms pins grease appropriately 5 Tighten I P mounting bolts 6 Refe...

Page 12: ......

Page 13: ......

Page 14: ...er user must therefore assume the ultimate responsibility for the proper sizing and selection installation operation and maintenance of Flowserve products The purchaser user should read and understand the INSERT OFFICIAL USER INSTRUCTION TITLE instructions included with the product and train its employees and contractors in the safe use of Flowserve products in connection with the specific applica...

Reviews: