ERPN-M USER INSTRUCTIONS ENGLISH 02-08
Page 6 of 37
run-up process (See section 5
Commissioning,
startup,
operation and shutdown).
NEVER RUN THE PUMP DRY
INLET VALVES TO BE FULLY
OPEN WHEN PUMP IS RUNNING
Running the pump at zero flow or below the
recommended minimum flow continuously will cause
damage to the magnets and/or bearings.
INLET VALVES TO BE FULLY
OPEN WHEN PUMP IS RUNNING
Running the pump at zero flow or below the
recommended minimum flow continuously will cause
damage to the seal.
DO NOT RUN THE PUMP AT
ABNORMALLY HIGH OR LOW FLOW RATES
Operating at a flow rate higher than normal or at a
flow rate with no back pressure on the pump may
overload the motor and cause cavitation. Low flow
rates may cause a reduction in
pump/bearing life,
overheating of the pump, instability and
cavitation/vibration.
When ambient temperatures are
likely to drop below freezing point, the pump and any
cooling and flushing arrangements must be drained
or otherwise protected.
HIGH MAGNETIC FIELDS
Persons working with permanent parts must be made
aware that there is a potential danger from magnetic
fields e.g. influence on heart pacemakers, credit
cards, computers, computer tapes/discs and
watches.
Great care must be taken when working with or
assembling/ dismantling magnetic rotors because of
the very high forces which can be created by the
magnets.
When shipping raw or bare magnet assemblies,
especially by air, special precautions may be
necessary.
HANDLING COMPONENTS
Many precision parts have sharp corners and the
wearing of appropriate safety gloves and equipment
is required when handling these components. To lift
heavy pieces above 25 kg (55 lbs) use a crane
corresponding to the mass and in accordance with
current local regulations.
NEVER DO MAINTENANCE WORK WHILST THE
UNIT IS CONNECTED TO POWER
HAZARDOUS LIQUIDS
When the pump is handling hazardous liquids care
must be taken to avoid exposure to the liquid by
appropriate sitting of the pump, limiting personnel
access and by operator training. If the liquid is
flammable and/or explosive strict safety procedures
must be applied.
Gland Packing must not be used when pumping
hazardous liquids.
DRAIN PUMP AND ISOLATE PIPEWORK
BEFORE DISMANTLING THE PUMP
The appropriate safety precautions should be taken
where the pumped liquids are hazardous.
FLUORO-ELASTOMERS (When fitted)
When a pump has experienced temperatures over
250 °C (482 ºF), partial decomposition of fluoro-
elastomers (example: Viton) will occur. In this
condition these are extremely dangerous and skin
contact must be avoided.
GUARDS MUST NOT BE REMOVED WHILE
PUMP IS OPERATIONAL
THERMAL SHOCK
Rapid changes in the temperature of the liquid within
the pump can cause thermal shock, which can result
in damage or breakage of components and should be
avoided.
NEVER APPLY HEAT TO REMOVE
IMPELLER
Trapped lubricant or vapour could cause an
explosion.
HOT AND COLD PARTS
If hot or freezing components or auxiliary heating
supplies can present a danger to operators, they
must
be shielded to avoid accidental contact. If
complete protection is not possible, the machine
access must be limited to maintenance staff only.
Note: bearing housings must not be insulated and
drive motors and bearings may be hot.
If the temperature is greater than 68 °C (175 °F) or
below 5 °C (20 °F) in a restricted zone, or exceeds
local regulations, action as above shall be taken.
1.6.4 Products used in potentially explosive
atmospheres
Measures are required to:
•
Avoid excess temperature