1528 Final position control: The
condenser cooling water valve is
not open
No start ENABLE or emergency stop
Cooling water valve condenser:
The valve is actuated and the open
NAMUR
Control the end switch and the valve. Control
the compressed air supply
Operator
1529 Final position control: The
condenser cooling water valve is
not closed
No start ENABLE or emergency stop
Cooling water valve condenser:
The booster valve is not actuated and
the closed NAMUR sensor in no final
Control the end switch and the valve. Control
the compressed air supply
Operator
1530 Final position control:
The booster valve is not open
No start ENABLE or emergency stop
The booster valve is actuated and the open
NAMUR sensor registers no final position.
Control the end switch and the booster valve.
Control the compressed air supply.
Operator
1531 Final position control:
The booster valve is not closed
No start ENABLE or emergency stop
The booster valve is not actuated and
the closed NAMUR sensor in no final
Control the end switch and the booster valve.
Control the compressed air supply.
Operator