background image

Flowserve Corporation, Valtek Control Products, Tel. USA 801 489 8611

48-7

rate calibration equipment of 

±

0.1 percent of

span should be used. The pressure gauge port is

1

/8-inch NPT. Calibration manifolds are available

from the factory (Part No. 97370).

1. Connect I/P module to a supply pressure between

30 to 150 psi.

2. Remove I/P module housing cover. (See Figure 9.)

WARNING: Be certain power to I/P module is
disconnected before removing housing cover in
explosive atmospheres; otherwise personal in-
jury may occur.

3. Before adjusting the zero and span, be certain the

MPC feature is disabled. Refer to Step 7 in the
‘Adjusting the Minimum Pressure Cutoff Feature’
section.

4. Connect a current source to terminal block on

circuit board.

NOTE: The zero and span adjustments are multi-
turn potentiometers (pots) and have no stops on
the ends of their travel; however, they have a slip
clutch to prevent damage from over-adjustment.
The pots also make a clicking noise when they
have reached adjustment limits.

5. Apply a 4.0 mA signal to the input. Locate and

adjust zero trim pot to achieve a 3.0 psi output. The
output will increase with clockwise rotation of the
zero trim pot. If calibrating an I/P module with a 10-
50 mA input signal, apply a 10.0 mA signal to input.

6. Increase input signal to 20.0 mA (50 mA for  10-50

mA units). Locate and adjust span trim pot to
achieve a 15.0 psi output. The output will increase
with clockwise rotation of the span.

7. Recheck zero setting by repeating Step 5. The span

adjustment may affect zero setting.

8. Repeat Steps 5, 6 and 7 until proper adjustments

are obtained.

Adjusting Minimum Pressure Cutoff
Feature

The XL90 positioner with I/P Transducer has a ‘Mini-
mum Pressure Cutoff’ (MPC) feature, which allows the
user to set the positioner. When the input signal falls
below a user-adjustable current, the pressure output
falls rapidly to approximately 1.7 psi, causing the valve
to move to the failure position. This feature is generally
used when the service requires a tight shut off or to
prevent throttling near the valve seat. To adjust this
feature, refer to Figure 9 and perform the following:

NOTE: The following procedure applies only when
the minimum pressure cutoff feature will be used.

NOTE: The zero and span settings of both the
positioner and I/P transducer should be verified
as  accurate before the minimum pressure cutoff
feature is enabled and adjusted.

1. Connect the I/P module to a 30 to 150 psi air supply

pressure.

2. Remove the I/P module housing cover.

WARNING: Be certain power to the I/P module is
disconnected before removing the housing
cover in explosive atmospheres; otherwise per-
sonal injury may occur.

3. Connect an adjustable current source to the termi-

nal block on the circuit board. Apply the desired
input signal to the positioner at which the output
pressure is to fall to approximately 1.7 psi. This
signal can range from factory setting of 3.7 to 8 mA.

4. Turn the minimum pressure cutoff pot clockwise

until the output pressure drops off.

5. Fine-tune the pressure drop-off point by increasing

the input signal and then decreasing it through the
desired shut-off signal. Observe the signal value at
which the pressure drops off. If the pressure drops
off at a lower mA signal than desired, turn the MPC
pot counterclockwise. If the pressure drops off at a
higher signal than desired, turn the tight shut-off
screw clockwise.

6. Repeat Step 5 until the pressure drops off at the

desired input signal.

7. To disable the MPC feature turn the minimum

pressure cutoff pot (marked “MPC”) 20 turns  coun-
terclockwise or until it makes a clicking noise.

Positioner Maintenance

NOTE: Refer to NT 3000 IOM for I/P module
maintenance instructions.

For proper maintenance, proceed as follows:

1. Maintain a clean air supply, free of dust, oil and

water. A coalescing air filter for I/P is required to
ensure a clean air supply. Check and maintain filter
regularly.

2. Make sure all arms and levers move freely.

3. Check for loose parts.

4. Be sure there are no leaks in the air supply tubing,

fittings or connections.

5. Refer to the troubleshooting chart on page 12 in

case of problems.

Pilot Relay Disassembly and Reassembly

The pilot relay is available as a complete unit and can
be easily replaced (See Steps 2 and 18.) Before at-
tempting to correct any problem with the pilot relay
assembly, obtain a positioner repair kit that contains the
soft goods most commonly required.

NOTE: Numbers in parentheses correspond to
the numbers in Figure 16.

1. Remove the feedback spring (47) and rotate the

span and zero arms (40, 46) out of the way.

Summary of Contents for XL90 Series

Page 1: ...use the I P NT 3000 Transducer The I P 2000 Trans ducer is not acceptable for use with the XL90 Series Positioner The XL90 positioner can handle supply pressures up to 150 psi thus a supply regulator is usually not required However a five micron air filter is required for pneu matic positioners and a coalescing filter is required for I P positioners NOTE The air supply should conform to ISA Standa...

Page 2: ...the capsule moves up it will close the exhaust seat of the upper pilot poppet and open the supply seat which applies increased air pressure to the bottom cylinder port At the same time the pilot valve capsule will open the exhaust seat for the lower pilot poppet thus decreasing pressure to the top cylinder port This difference in pressure will drive the piston outward rotate the pinion and stretch...

Page 3: ...rm lockwasher and nut on the end of the shaft and tighten nut Carefully hold the shaft with wrench on shaft flats 4 With valve in closed position make sure follower pin lever is mounted parallel to valve lever flush with end of valve shaft and the nut is tight 5 Place follower pin mounted to the lever that is mounted to the valve shaft into follower arm slot and bolt the transfer case plate in pla...

Page 4: ... With a 50 percent input signal tighten all mounting bolts Stroke actuator to verify proper alignment POSITIONER CALIBRATION Introduction Valtek positioners are calibrated at the factory how ever due to shipping and handling it may be necessary to check the calibration before operating the valve The XL90 positioner can be calibrated to a range of 3 15 two way split range 3 9 or 9 15 and three way ...

Page 5: ...tmentscrew ment knob until valve begins to stroke with more than 3 psi signal for 9 15 psi range adjust to 9 psi 2 Loosen range adjustment locking screw no more than 1 8 turn 3 With a phillips screwdriver adjust the span adjust ment so valve is at full stroke with more than 15 psi for 3 15 or 9 15 psi range adjust to 9 psi for 3 9 psi range 4 Return to 3 psi or 9 psi for 9 15 psi range and check t...

Page 6: ...e 8 4 By grasping adjust lever carefully rotate gain ad just assembly to desired position CAUTION To avoid damaging gain adjust con necting spring mechanism Make sure both up per and lower gain adjust plates rotate together When they are rotated to the new position the connectingspringshouldbeperpendiculartothe plates 5 When the gain is set to the desired position firmly tighten both lock down scr...

Page 7: ...pressurecutofffeaturewillbeused NOTE The zero and span settings of both the positioner and I P transducer should be verified as accurate before the minimum pressure cutoff feature is enabled and adjusted 1 Connect the I P module to a 30 to 150 psi air supply pressure 2 Remove the I P module housing cover WARNING Be certain power to the I P module is disconnected before removing the housing coverin...

Page 8: ...seat is removed by pushing it out with a soft instrument such as a wooden dowel Be careful not to damage seating surface Remove O ring 37 from seat ring 13 Lubricate and replace O ring 37 on movable seat ring 34 Carefully reinstall seat ring into upper relay body 9 being careful not to damage seating surface or O rings 14 Reinstall poppets 28 poppet springs 29 Seat Spring O rings 30 and poppet cov...

Page 9: ... and check three screws holding positioner shifts to bracket check two bolts holding bracket to yoke 2 Loose linkage 2 Tighten nuts and bolts on linkage and stem clamp 3 Loose zero adjustment locking knobs 3 Tighten zero adjustment locking knob re calibrate if necessary 4 Wear of arms or pins 4 Replace worn arms pins grease appropriately 5 I P mounting bolts loose 5 Tighten I P mounting bolts 6 St...

Page 10: ...justable gain upper plate 22 Pan head screw 23 Socket screw 24 Washer 25 Spacer nut 26 Adjustable gain lower plate 27 Poppet cover 28 Poppet 29 Poppet spring 30 O ring 31 Retaining ring 32 Screw 33 Screw 34 Adjustable seat 35 Rubber cap 36 Balance adjust screw 37 O ring 38 O ring 39 Adjustable seat spring 40 Span arm 43 Pivot bushing 44 Pivot screw 45 Snap ring 46 Zero arm 47 Feedback spring 48 Pi...

Page 11: ...sembly 10 Orifice screw 11 O ring 12 O ring 13 Relay diaphragm assembly 14 Relay plate 15 Diaphragm retaining plate 16 Signal diaphragm assembly 17 Set screw 18 Spring 19 Set screw 20 Nut 21 Adjustable gain upper plate 22 Pan head screw 23 Socket screw 24 Washer 25 Spacer nut 26 Adjustable gain lower plate 27 Poppet cover 28 Poppet 29 Poppet spring 30 O ring 31 Retaining ring 32 Screw 33 Screw 34 ...

Page 12: ...m a C r a e n i L e e r g e D 0 9 0 6 5 4 0 3 d r o c c A C m a C r e p m a D t o o R e r a u q S e r a u q S r a e n i L d r o c c A D s e v l a V y l f r e t t u B m a C e e r g e D 0 6 0 d r o c c A E r o t a u t c A m g a r h p a i D m a C d n a r a e n i L g e r e S F r o t a u t c A n o t s i P m a C r a e n i L k e t l a V D A E R H T T I U D N O C N O I T C E N N O C P d r a d n a t S T P ...

Page 13: ... 0 1 o N t r a P t i K y a l e R p m e T t x E o N m e t I n o i t p i r c s e D y t i t n a u Q 9 6 s e d u l c n I y l b m e s s a y a l e R 0 7 7 5 6 5 9 3 8 s o N m e t i 7 7 6 7 4 7 3 7 2 7 1 8 g n i r O 4 3 3 w e r c s d a e h n a P 4 7 7 5 8 7 0 0 1 o N t r a P t i K g n i r O p m e T t x E o N m e t I n o i t p i r c s e D y t i t n a u Q 8 d n a g n i r O e s a b y a l e R g n i r O e b u...

Page 14: ...e k c i t S 1 g e r e S e t i h W s n e L t i K r e v o C 5 9 8 5 1 1 0 1 o N t r a P o N m e t I n o i t p i r c s e D y t i t n a u Q 2 r e v o C 1 3 w e r c S 4 7 1 1 g n i r O 1 9 1 1 s n e L g e r e S 1 6 1 1 r o t a c i d n I 1 8 1 1 g n i r O 1 d r o c c A k c a l B s n e L t i K r e v o C 8 9 8 5 1 1 0 1 o N t r a P o N m e t I n o i t p i r c s e D y t i t n a u Q 2 r e v o C 1 3 w e r c ...

Page 15: ... t c e l E t i u c r i c g n i r p s s m a c t f a h s g n i t n u o m d n a d r a o b s w e r c s d n a s r e c a p s 1 d n E t f a h S R U M A N t i K h c t i w S t i m i L l a c i n a h c e M 0 1 9 5 1 1 0 1 o N t r a P n o i t c e n n o C o N m e t I n o i t p i r c s e D y t i t n a u Q 2 5 t f a h S m a C 1 2 1 1 g n i R g n i n i a t e R 2 8 2 1 t u N m a J 1 0 3 1 s e d u l c n I t i K s c...

Page 16: ...ntact www flowserve com or call USA 972 443 6500 Regional Headquarters 1350 N Mt Springs Prkwy Springville UT 84663 Phone 801 489 8611 Facsimile 801 489 3719 12 Tuas Avenue 20 Republic of Singapore 638824 Phone 65 862 3332 Facsimile 65 862 4940 12 av du Québec B P 645 91965 Courtaboeuf Cedex France Phone 33 1 60 92 32 51 Facsimile 33 1 60 92 32 99 Quick Response Centers 5114 Railroad Street Deer P...

Reviews: