APPENDIX C: COMMUNICATION VARIABLES
Remarks:
Below, an overview of the F170-A-OS specific variables; other common variables are described
in the standard table.
All numbers are decimal numbers, unless otherwise noted.
Following variables of the standard table (var00-var30) are not valid for this product and will be
responded with value 1: var00, 03-05, 07,08, 16-22, 24, 26-29.
CONFIGURATION VARIABLES F170-A-OS - SETUP-LEVEL:
VAR DESCRIPTION
BYTES VALUE
REMARKS
LEVEL
48
(30h)
unit 1
0=mL
1=L
2=m3
3=mg
4=g
5=kg
6=ton
7=gal
8=bbl
9=lb
10=cf
11=rev
(revolutions for
RPM)
12=none
50
(32h)
decimals 1
0…1
51
(33h)
span
3
1....9.999.999
S 0000001 up to S 0000009 is
allowed when decs < 6! (VAR54)
54
(36h)
decimals span
1
0…6
ALARMS
234
EAh
level low
3
0-9,9999
decimals: see 50 (32h)
237
EDh
level high
3
0-9,9999
decimals: see 50 (32h)
205
CDh
delay time alarm
low level
2
1…9,999
steps of 0.1 second
DDh
delay time alarm
high level
2
1…9,999
steps of 0.1 second
44h
edit level alarm
1
0=operator
1=SETUP level
46h
alarm at level zero
1
0=ignore
1=default
2=no relay
DISPLAY
68
(44h)
set level monitor
1
0=operator level
1=SETUP level
POWERMANAGEMENT
80
(50h)
LCD update time
1
0=fast
1=1sec
2=3sec
3=15sec
4=30sec
5=off
81
(51h)
power-mode battery 1
0=operational
1=shelf