ESA614
Getting Started Manual
14
For More Information
For more information on how to use the Product, refer to
the
ESA614 Users Manual
at www.flukebiomedical.com.
Maintenance
XW
Warning
To prevent possible electrical shock, fire, or
personal injury:
•
Turn off the Product and remove the
mains power cord. Stop for 2 minutes to
let the internal circuits discharge before
you open the fuse door or remove
Product covers.
•
Do not operate the Product with covers
removed or the case open. Hazardous
voltage exposure is possible.
•
Disconnect the mains power cord before
you remove the Product covers.
•
Remove the input signals before you
clean the Product.
•
Use only specified replacement parts.
•
Use only specified replacement fuses.
•
Have an approved technician repair the
Product.
The Product is a calibrated measurement instrument. Use
the necessary precautions to prevent mechanical abuse
that could change the calibrated adjustments.
Fuse Test and Fuse Replacement
XW
Warning
To prevent electric shock, remove all power
cords and test leads from the Product before
opening the fuse door.
For electrical protection of the equipment outlet, the
Product uses two fuses, one in the live (L1) line and one
in the neutral (L2) line. To do a fuse test:
1. Turn the Product so the case bottom is up. See
Figure 6.
2. Flip up the tilt stand.
3. Remove the screw in the fuse door with a #2 Phillips
head screwdriver and lift the fuse door from the
Product.
4. Remove the two fuses from the Product.