Electrical Safety Analyzer
Maintenance
15
F1 - F2
gtv111.eps
Figure 5. Fuse Access
5. Using a multimeter, measure the continuity of each
fuse.
If a fuse does not show continuity, replace the fuse(s)
with a fuse of the same current and voltage rating.
Applicable fuse ratings are posted on the case
bottom label of the Product. Table 5 is a list of
available fuses with Fluke Biomedical part numbers.
6. Reinstall the fuse door and secure it with the screw.
How to Clean the Product
XW
Warning
To prevent electric shock, do not clean the
Product plugged into mains or attached to a
DUT.
W
Caution
Do not pour fluid onto the Product surface.
Fluid in the electrical circuitry can cause the
Product to fail.
W
Caution
Do not use spray cleaners on the Product.
This can cause fluid to leak into the Product
and damage electronic components.
Clean the Product occasionally with a damp cloth and
weak detergent. Try to prevent the entrance of liquids.
Clean the adapter cables with the same precautions.
Examine them for damage and deterioration of the
insulation. Examine the connections for integrity before
each use.