Electrical Safety Analyzer
Specifications
Specifications
Temperature
Operating ............................................................ 10
°
C to 40
°
C (50
°
F to 104
°
F)
Storage ................................................................ -20
°
C to 60
°
C (-4
°
F to 140
°
F)
Humidity
................................................................. 10 % to 90 % non-condensing
Altitude
................................................................... To 5,000 meters @ 115 V ac mains and
≤
150 V measurements
To 2,000 meters @ 230 V ac mains and
≤
300 V measurements
Display
.................................................................... LCD display
Communications
................................................... USB device port for computer control
Modes of Operation
............................................... Manual and remote
Power
120 Volt power outlet .......................................... 90 to 132 V ac rms, 47 to 63 Hz, 20 A maximum
230 Volt power outlet .......................................... 180 to 264 V ac rms, 47 to 63 Hz, 16 A maximum
Size (L x W x H)
...................................................... 32 cm x 23.6 cm x 12.7 cm (12.6 in x 9.3 in x 5 in)
Weight
.................................................................... 4.7 kg (10.25 lb)
Safety
...................................................................... IEC 61010-1: Overvoltage category II, Pollution Degree 2
IEC 61010-2-030: Measurement 300 V, CAT II
Electromagnetic Compatibility (EMC)
International
....................................................... IEC 61326-1: Controlled Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled
radio frequency energy that is necessary for the internal function of the equipment
itself.
17
Summary of Contents for ESA620
Page 6: ......
Page 8: ...ESA620 Getting Started Manual Specifications 17 Detailed Specifications 18 ii ...
Page 10: ...ESA620 Getting Started Manual iv ...
Page 12: ...ESA620 Getting Started Manual vi ...
Page 22: ...ESA620 Getting Started Manual faw03 eps Figure 3 DUT Connections to the Analyzer 10 ...