ESA620
Getting Started Manual
Equipment Current
Range ................................................................. 0 – 20 A ac rms
Accuracy ............................................................. 5 % of reading
±
(2 counts or 0.2 A, whichever is greater)
Duty cycle ........................................................... 15 A to 20 A, 5 min. on/5 min. off
10 A to 15 A, 7 min. on/3 min. off
0 A to 10 A continuous
Leakage Current
Modes* ................................................................ AC+DC (True-rms)
AC only
DC only
* Modes: AC+DC, AC only, and DC only available for all leakages with exception of MAP
that are available in True RMS (shown as AC+DC)
Patient Load Selection ........................................ AAMI ES1-1993 Fig. 1
IEC 60601: Fig 15
IEC 61010: Fig A-1
Crest factor .........................................................
≤
3
Ranges ................................................................ 0.0 to 199.9
µ
A
200 to 1999
µ
A
2.00 to 10.00 mA
Accuracy**
DC to 1 kHz .....................................................
±
(1 % of r (1
µ
A or 1 LSD, whichever is greater))
1 to 100 kHz ....................................................
±
(2 % of r (1
µ
A or 1 LSD, whichever is greater))
100 kHz to 1 MHz ...........................................
±
(5 % of r (1
µ
A or 1 LSD, whichever is greater))
** Map Voltage: Additional residual leakage up to 4
µ
A @ 120 V ac, 8
µ
A @ 240 V ac
20