KIT-EMM300 

Electrical Measurement Module Kit with Docking Station

 

 

Instructions 

Introduction  

The Electrical Measurement Module with Docking Station (the Product) is an accessory to allow electrical 
measurements to be made from pressure Devices Under Test (DUT) such as transmitters, transducers, and more. 
The Product includes an EMM300 Electrical Measurement Module and its dock, the DS70-KIT-EMM Docking Station 
that enables connection of the Electrical Measurement Module (EMM) to a pressure calibrator. The EMM module 
measures mA, V dc, and can supply 24 V loop power. It also provides basic analog trimming of HART devices. 

How to Contact Fluke Calibration 

To contact Fluke Calibration, call one of the following telephone numbers: 

  Technical Support USA: 1-877-355-3225 

  Calibration/Repair USA: 1-877-355-3225 

  Canada: 1-800-36-FLUKE (1-800-363-5853) 

  Europe: +31-40-2675-200 

  Japan: +81-3-6714-3114 

  Singapore: +65-6799-5566 

  China:  +86-400-810-3435 

  Brazil:  +55-11-3759-7600 

  Anywhere in the world: +1-425-446-6110 

 

To see product information and download the latest manual supplements, visit Fluke Calibration’s website at 

www.flukecal.com

To register your product, visit 

http://flukecal.com/register-product

.

 

Summary of Contents for KIT-EMM300

Page 1: ...s and more The Product includes an EMM300 Electrical Measurement Module and its dock the DS70 KIT EMM Docking Station that enables connection of the Electrical Measurement Module EMM to a pressure calibrator The EMM module measures mA V dc and can supply 24 V loop power It also provides basic analog trimming of HART devices How to Contact Fluke Calibration To contact Fluke Calibration call one of ...

Page 2: ...ls The symbols shown in Table 1 are found in these instructions Table 1 Symbols Symbol Definition W WARNING RISK OF DANGER Consult user documentation Certified by CSA Group to North American safety standards Conforms to relevant Australian Safety and EMC standards P Conforms to European Union directives Earth This product complies with the WEEE Directive marking requirements The affixed label indi...

Page 3: ...tion The Product 3 The Product Table 2 shows the Product Table 2 The Product 1 2 4 5 6 3 ifh01 eps Number Description Docking Station DS70 KIT EMM Electrical measurement Module EMM300 Input Terminals D Power Socket E RS232 Connector F CAN bus Connector ...

Page 4: ...cator shows on the lower half of the main screen on the Pressure Controller 5 Connect the RS232 cable to your computer optional 6 Connect the DUT pressure connection s to the Pressure Controller TEST port or to an accessory connected to the TEST port CPS SPLT Test Station for example 015504001 100 240 VAC 47 63 Hz 100W MAX T2A 250V SERIAL NUMBER FLUKE CORPORATION 6920 SEAWAY BLVD M S 143F EVERETT ...

Page 5: ...troller s Operators Manual for how to use the Programs Task with mA or V dc transmitters Pressure Transmitter Instructions The Product tests troubleshoots and calibrates analog DUTs and smart pressure transmitters with HART functionality Pressure transmitters are usually classified as analog transmitters or smart transmitters Analog Transmitters Analog transmitters let the user make two adjustment...

Page 6: ...d when calibrated To communicate with smart transmitters HART communication protocol is used Many of the HART variables are used to calibrate the smart transmitter The EMM is HART enabled and uses Universal Practice Commands and many HART Common Practice Commands to change parameters and make adjustments to the HART device Many smart transmitters use specific commands that are not part of the comm...

Page 7: ...Electrical Measurement Module Kit with Docking Station Pressure Transmitter Instructions 7 iIfh02 eps Figure 2 Transmitter Connection ...

Page 8: ...g Tag Message Description 3rdVar 4thVar LRV URV Expanded Device Type Rev Dev Rev SW Rev HW Rev Device ID Number Of Variables Change Counter Critical Power Variable Alert Maintenance Required Manufacture Private Label Message Tag PV Unit PV PVAO PV 2nd Var Description Date Serial Number Address LTL UTL Min Span PV Alarm PV Damp Write Protect Transfer Function Preambles Simulation NVMEM Fault RAM Fa...

Page 9: ...abled when measuring value is stable 6 Push Send to send the 4 mA trimming value to DUT 7 After 4 mA trimming is complete push Trim 4mA to repeat the 4 mA trimming if necessary Or push Continue to move trim to 20 mA 8 Push Trim 20mA to start trimming the 20 mA point The Send button is enabled when measuring value is stable 9 Push Send to send 20 mA trimming value to DUT 10 After 20 mA trimming is ...

Page 10: ...rement is taken and the controller continues to the next setpoint If closed loop is OFF push continue to go to the next setpoint 3 In the Trim by Re ranging main menu push Start to start trim 4 If the Closed Loop was ON the trim process automatically progresses without user interaction Otherwise the Closed Loop is OFF 5 Push Trim to trim and change to next range 6 If necessary push Lower Range or ...

Page 11: ...lts if any are shown if any have occurred Write PV Unit The Write PV Unit command changes the Primary Variable pressure unit of the transmitter Note This can also be changed in the HART Data menu 1 Push HART Tasks on the HART menu 2 Select Write PV Unit 3 Select the new unit 4 Push Send 5 Push Exit Write Tag Write the short tag maximum 8 characters as follows Note This can also be changed in the H...

Page 12: ... the complete data about the transmitter model hardware and software revision numbers and many coefficients From the controller screen push the HART Data button to access the HART Information screen Some data is read only and cannot be selected Some data is writable see Figure 3 To view and change HART data 1 Push HART Data on the HART menu 2 Use the navigation keys to navigate between fields Only...

Page 13: ...95 relative humidity non condensing Weight 1080 g 2 4 lb Dimensions 165 x 240 x 85 mm 6 5 x 9 5 x 3 4 in Safety General IEC 61010 1 LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for one year from the date of purchase This warranty does not cover fuses disposable batteries or damage from accident neglect misuse alteration conta...

Reviews: