background image

 

September 2012 

© 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. 

All product names are trademarks of their respective companies. 

2700G Series 

Reference Pressure Gauge 

 
 
 

Users Manual 

 

1.888.610.7664 

sales@GlobalTestSupply.com

Fluke-Direct

.com

Summary of Contents for 095969376969

Page 1: ...rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 2700G Series Reference Pressure Gauge Users Manual 1 888 610 7664 sa...

Page 2: ...r replacement of a defective product which is returned to a Fluke authorized service center within the warranty period uke assumes no risk for damage in transit Following warranty repair the product w...

Page 3: ...pment 2 Safety Information 2 Special Conditions for Safe Use 3 Symbols 4 Display and Buttons 5 Operation 6 How to Setup the Product 6 Engineering Units 6 Set Auto Off 7 Show Battery Voltage 7 Display...

Page 4: ...roduct 9 How to Change the Batteries 10 Accessories 11 USB Interface Cable 11 Power Module 11 Serial Interface Instructions 13 Initiating Communication 13 List of Commands 13 Parameter Units 14 Error...

Page 5: ...iii List of Tables Table Title Page 1 Symbols 4 2 Display and Buttons 5 3 Units Conversion 16 1 888 610 7664 sales GlobalTestSupply com Fluke Direct com...

Page 6: ...v List of Figures Figure Title Page 1 The Product 5 2 Change the Batteries 10 3 USB Power Module and Universal Connectors 12 1 888 610 7664 sales GlobalTestSupply com Fluke Direct com...

Page 7: ...or in applications where high accuracy pressure measurement is required The Product features user configurable functions that include Sampling rate Tare Damping Auto off Min Max When the Product is co...

Page 8: ...perate high pressure systems if you know the correct safety procedures High pressure liquids and gases are hazardous and the energy from them can be released without warning To prevent possible electr...

Page 9: ...pecified Maximum torque specified is 20 Nm 15 ft lb Special Conditions for Safe Use Misuse If the Product is exposed to overpressure or sudden physical shock such as being dropped examine it for damag...

Page 10: ...Australian standards This product complies with the WEEE Directive 2002 96 EC marking requirements The affixed label indicates that you must not discard this electrical electronic product in domestic...

Page 11: ...s the display In Configure Mode push the button to move forward through the menus MIN MAX records minimum and maximum pressure values and saves them to memory Push to show maximum MAX indication Push...

Page 12: ...re value the pressure shown is not the actual pressure applied How to Setup the Product Before you use the product it is necessary to configure it for your application Push to go to the Setup menu Eac...

Page 13: ...e on and off Damping smooths readings from pulsating pressure sources Set Sample Rate This function finds how many times pressure is sampled and the display is updated Selections are 0 5 1 3 and 10 sa...

Page 14: ...ry Hold or down to move faster through the selections by a factor of 10 When you stop the counter push and again to move forward or backward by a factor of 1 The password is factory set and cannot be...

Page 15: ...with the backlight off When the battery voltage is low the low battery icon shows on the top left of the display To replace the batteries see the How to Change the Batteries section Maintenance How to...

Page 16: ...s see Figure 2 1 Pull off the Product cover 2 Use a Phillips screwdriver to loosen the captive screw on the battery door 3 Remove the battery door 4 Replace the three AA batteries 5 Install the batter...

Page 17: ...he back of the Product You can use serial communication to configure and calibrate the Product and move measurement data from the Product to a PC For specifications on the interface see the Specificat...

Page 18: ...2700G Series Users Manual 12 gsn003 eps Figure 3 USB Power Module and Universal Connectors 1 888 610 7664 sales GlobalTestSupply com Fluke Direct com...

Page 19: ...of commands to communicate with the Gauge CLS Clears the error queue FAULT Returns an error code from the error queue IDN Identification query Returns the manufacturer model number and firmware revis...

Page 20: ...its The subsequent list of units is used with the Product psi Pressure in pounds per square inch bar Pressure in bars mBar Pressure in millibars kg cm2 Pressure in kilograms per centimeter squared inH...

Page 21: ...uge uses the subsequent error codes 101 A non numeric entry was received where it should be a numeric entry 102 Too many significant digits entered 103 Invalid units or parameter value received 105 En...

Page 22: ...kPa Conversion Factor to kPa psi 0 14503773773 6 894759 bar 0 01 100 MPa 0 001 1000 kgf cm2 0 010197162130 98 06652 inH2O 4 C 4 014742 0 249082 inH2O 20 C 4 021845 0 248642 inH2O 60 F 4 0185886 0 248...

Page 23: ...o kPa cmH2O 4 C 10 19744 0 09806383 cmH2O 20 C 10 21549 0 09789056 mH2O 4 C 0 1019744 9 806383 mH2O 20 C 0 1021549 9 789056 kPa 1 1 mbar 10 0 1 inHg 0 C 0 2952998 3 386387 mmHg 0 C 7 500618 0 133322 T...

Page 24: ...on corrosive gas 100 300 500 1000 psi any liquids or gases compatible with 316 stainless steel Above 1000 psi any non flammable non toxic non explosive non oxidizing liquid or gas compatible with 316...

Page 25: ...n depth 1 5 in Without Protective Cover Pressure Connection in NPT Male Housing Cast ZNAL Display 5 1 2 Digits 16 53 mm 0 65 in high 20 Segment bar graph 0 to 100 Power Battery three size AA alkaline...

Page 26: ...0 10000 Pressure Range MPa 0 1 0 2 0 7 2 3 5 7 20 35 70 Vacuum Range psi 15 15 12 12 12 12 0 0 0 Vacuum Range kPa 100 100 80 80 80 80 0 0 0 Burst Pressure psi 45 90 1000 2000 2000 10000 10000 10000 15...