TiS55+/TiS75+
Users Manual
12
Measurement Menu
Table 8
is a list of the options available in the Measurement menu.
Table 8. Measurement Menu
Option Menu
Option
Description
Level/Span
Auto
Sets the Level/Span to adjust automatically or
manually.
Manual
Set Level/Span
With Level/Span set to
Manual
, changes the
Level/Span. See
Emissivity
Custom Value
Sets a custom emissivity value when a value
from the standard emissivity table is not
appropriate for the measurement. See
<options>
Select an emissivity value from a list of common
materials. Use the scroll bar to see all of the
options. See
.
Background
<options>
Changes the background temperature to
compensate for reflected background
temperature.
Very hot objects or very cold objects can affect
the apparent temperature and measurement
accuracy of the target, especially when surface
emissivity is low. Adjust the reflected
background temperature to improve the
accuracy of the measurement.
Transmissivity
<options>
Changes the values used to calculate the
temperature based on the transmission
percentage of the infrared-transparent window
(IR window). See
.
www.
.com
information@itm.com