Error Messages and Troubleshooting
Error Messages
7
7-15
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 651
USB memory not
ready
Cause:
USB memory is not ready for data record.
Solution:
You may have configured the record destination to
USB, but the memory is not ready when recording is started.
Error 652
Data file limit
exceeded
Cause:
Too many data files were created for a scan.
Solution:
Start a new scan.
Error 653
Error creating record
file
Cause:
Failed to create file for data record.
Solution:
Make sure that the destination memory is not full.
Error 654
Error writing record
file
Cause:
Failed to write file for data record.
Solution:
Make sure that the destination memory is not full.
Error 655
Error resuming data
record
Cause:
Failed to resume the data record.
Solution:
Make sure that the destination memory is not full.
Error 656
Data record limit
exceeded
Cause:
Too many data records in internal memory
(Max.1000).
Solution:
Remove some data records.
Error 657
Incorrect timestamp
Cause:
RTC clock is reset.
Solution:
Set the instrument’s date time setting.
Error 658
Record state
damaged
Cause:
The battery could have failed.
Solution:
Replace the battery.
Error 700
NV memory invalid
Cause:
The flash memory contains invalid Instrument setup,
LAN setup, MAC setup or calibrate constants.
Solution:
Default settings will be loaded, if the MAC setup or
calibrate constants show failed, Contact Fluke. See Chapter
1.
Error 701
NV invalid so default
loaded
Cause:
Default settings have been loaded.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 702
Error reading
instrument setup
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 703
Error writing
instrument setup
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Summary of Contents for 1586A
Page 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Page 8: ...1586A Users Manual vi...
Page 10: ...1586A Users Manual viii...
Page 34: ...1586A Users Manual 2 2...
Page 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Page 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Page 46: ...1586A Users Manual 2 14...
Page 48: ...1586A Users Manual 3 2...
Page 82: ...1586A Users Manual 3 36...
Page 84: ...1586A Users Manual 4 2...
Page 108: ...1586A Users Manual 5 2...
Page 114: ...1586A Users Manual 6 2...
Page 118: ...1586A Users Manual 6 6...
Page 120: ...1586A Users Manual 7 2...