1586A
Users Manual
3-8
Table 3-1. Types of Inputs
Type of Input
Range and Types
Channel
Configuration
Reference
Wiring Polarity
DC Voltage
Range: 0 V to 50 V
Page 3-19
H
L
V
DC Current
[1]
Range: 0 mA to 100 mA
Page 3-19
H
L
V
Resistance (
Ω
)
2-Wire or 4-Wire
[3]
Range: 0
Ω
to 100 M
Ω
Page 3-20
H
L
R
2-Wire
H
Source
Sense
L
R
4-Wire
H
L
H
Source
Sense
L
R
3-Wire
H
L
Platinum
Resistance
Thermometer
(PRT)
2-Wire, 3-Wire or 4-Wire
Types: ITS90, CVD,
PT385, PT392
Page 3-23
Thermistor
2-Wire or 4-Wire
Types: R(T)
[2]
, 2.252
k
Ω
, 5 k
Ω
, 10 k
Ω
Page 3-22
Thermocouple
Types: B, C, D, E, G, J,
K, L, M, N, R, S, T, U,
W, POLY
[2]
Page 3-21
H
L
Note
[1]
−
Each Input Module has two terminal sets (mA 21 and mA 22) that are dedicated to current measurements.
[2] – Custom characterized.
[3]
−
100 M
Ω
is for Ch001 only. Other analog channels are 10 M
Ω
Max.
Input Wiring Instructions
Use the procedure below and refer to Figure 3-1 for instructions on how to wire a 2-wire,
3-wire, or 4-wire input to the Input Module.
Warning
To prevent possible electrical shock, fire, or personal injury,
read the Wiring Safety and Considerations section on page 3-5.
1.
Power off the Product with the main power switch.
Summary of Contents for 1586A
Page 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Page 8: ...1586A Users Manual vi...
Page 10: ...1586A Users Manual viii...
Page 34: ...1586A Users Manual 2 2...
Page 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Page 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Page 46: ...1586A Users Manual 2 14...
Page 48: ...1586A Users Manual 3 2...
Page 82: ...1586A Users Manual 3 36...
Page 84: ...1586A Users Manual 4 2...
Page 108: ...1586A Users Manual 5 2...
Page 114: ...1586A Users Manual 6 2...
Page 118: ...1586A Users Manual 6 6...
Page 120: ...1586A Users Manual 7 2...