Digital Multimeters
Performance Tests
5
lists the equipment required for the performance tests.
Before you do the performance test, make sure to warm up the calibrator.
To test each function and operating ranges:
1. For each step in
, set the DUT to the specified function and range.
2. Connect the source to the input jacks on the DUT.
3. Apply the output from the source. For temperature connections, see
4. Make sure the DUT readings are within the limits in
.
Table 1. Required Equipment
Equipment
Recommended Model
Calibrator
Fluke Calibration 5502A, 5502E, or 5522A
Type K to dual banana adapter
Fluke 80AK-A
Table 2. Performance Specifications
Step
Function
Calibrator Output
15B
MAX
17B
MAX
Lower
Limit
Upper
Limit
Value
Frequency or
Amplitude
1
VAC
5.5 V
50 Hz
√
√
5.442
5.558
2
55 V
50 Hz
√
√
54.42
55.58
3
55 V
500 Hz
√
√
54.42
55.58
4
550 V
50 Hz
√
√
544.2
555.8
5
1000 V
50 Hz
√
√
986
1014
6
VAC/Freq.
45 Hz
1 V
X
√
44.92
45.08
7
5 kHz
1 V
X
√
4.992
5.008
8
100 kHz
3 V
X
√
99.6
100.4
9
VAC/Duty
50 Hz
3 V (Square
Wave)
X
√
49.5
50.5
10
VDC
5.5 V
0 Hz
√
√
5.470
5.531
11
55 V
0 Hz
√
√
54.70
55.31
12
550 V
0 Hz
√
√
547.0
353.1
13
-1000 V
0 Hz
√
√
-1008
-992
14
mVDC
550 mV
0 Hz
√
√
543.5
556.5
15
mVAC
550 mV
500 Hz
√
√
533.2
566.8