Power Logger
Technical
Specifications
53
Operating error:
±
(0.5 % of m. v. + 10 digit)
Resolution: 0.1
V
V-RMS Delta measurement
Measuring range:
100 / 115 / 190 /208 / 220 / 380 / 400 / 415
/ 450 / 480 / 600 / 660 / 690 / 720 / 830 V
AC
Intrinsic error:
±
(0.2 % of m. v. + 5 digit)
Operating error:
±
(0.5 % of m. v. + 10 digit)
Resolution: 0.1
V
A-RMS measurement
Fluke Flexi sets and current probes with voltage output are supported. All
current probes must correspond to 600 V / CAT III
Flexi set I
N
ranges:
15 / 150 / 3000 A RMS (at sine)
Current clamp ranges:
5 / 50 / 500 mV AC
Resolution: 0.01
A
Ranges:
150 / 3000 A
50 mv / 500 mV output
Intrinsic
error:
±
(0.5% of m. v. + 10 digit)
Operating
error:
±
(1% of m. v.+ 10 digit)
15 A
5 mV output
Intrinsic
error:
±
(0.5% of m. v. + 20 digit)
Operating
error:
±
(1% of m. v. + 20 digit)
The errors of the current probes are not considered.
By using Fluke flexi-set:
Flexi set measuring error:
±
(2 % of m. v. + 10 digit)
Position influence:
±
(3 % of m. v. + 10 digit)
CF (typical):
2.83
Power measurement (P, Q, S, D)