67 MAX
Users Manual
10
Use the Product
Note
When ambient conditions change significantly, wait 30 minutes for the Product to
stabilize before you use it or the accuracy may be impacted. For example, if the
Product is brought from cold or hot outdoor conditions into a building.
Aim the Product for Body Temperature
When you use the Product to measure body temperature,
make sure
that the Product shows
the correct temperature. When you measure the subject’s temperature, make sure the skin of
the forehead is unobstructed by hair, eye glasses, or anything that could interfere with the
emitted infrared energy. Make sure the distance to subject is 50 mm to 150 mm (2 in to 6 in),
or accuracy may be impacted. Make sure the Product is in
Body
mode. See Figure 1.