Maintenance
Accuracy Performance Tests
4
4-23
Resistance Range*
Decade Resistor
Minimum Reading
Maximum Reading
30 k
Ω
Short Circuit (Zero)
700
Ω
1
k
Ω
30 k
Ω
29
k
Ω
29.681
k
Ω
30.019
k
Ω
300 k
Ω
290
k
Ω
289.07
k
Ω
292.63
k
Ω
3 M
Ω
2.9
M
Ω
2.8607
M
Ω
2.9410
M
Ω
*
The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire
resistance plus 0.01% decade resistance tolerance. You must add any additional lead wire
resistance present in your setup to the resistance values given in this table
.
Resistance Range*
5700A
Minimum Reading
Maximum Reading
30 k
Ω
Short Circuit (Zero)
700
Ω
1
k
Ω
30 k
Ω
19
k
Ω
19.686
k
Ω
20.014
k
Ω
300 k
Ω
190
k
Ω
189.60
k
Ω
192.10
k
Ω
3 M
Ω
1.9
M
Ω
1.8740
M
Ω
1.9277
M
Ω
*
The resistance accuracy in this table makes allowance for up to 0.1
Ω
of lead wire resistance.
You must add any additional lead wire resistance present in your setup to the resistance values
given in this table. Using the 5700A in 2-wire compensation mode eliminates much of the lead
resistance errors. See the 5700A Operator’s Manual for details
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...