2680A/2686A
Users Manual
A-2
2680A/2686A General Specifications
Table A-1 provides the general specifications for the 2680 Series instruments.
Table A-1. 2680 Series General Specifications
Specification Characteristic
Input Channel Capacity
Maximum of 120 channels (Precision or Fast Analog Input) per
chassis
Optional DIO
20 digital input and output channels, 8 double pole-single throw
relay channels, a totalizer input, and a totalizer enable input.
Computed Channels
60
Size
473 mm (18.6 in) x 432 mm (17 in) x 237 mm (9.3 in)
Weight
2680A/2686A (empty)
8.6 Kg (18.9 lb)
2680A
–
FAI 0.8
Kg
(1.8
lb)
2680A
–
PAI 1.2
Kg
(2.7
lb)
2680A
–
DIO 0.8
Kg
(1.8
lb)
Power
100 – 240 V ac (no switching required), 50 to 60 Hz, 100 VA
maximum or optional 9 V dc to 45 V dc, 35 W maximum
EMC EN50082-2
EN55022-1
EN55011 class A
EN610000-4-2,3,4,6,8
EN61326
Safety
EN61010-1, CAT II (DIO is rated CAT I)
CSA C22.2 No. 1010.1
Serial Interface
Connector: 9 pin male (DB9)
Signals: Tx, Rx, DTR, RTS, GND
Modem Control: full duplex
Baud rates: 4800, 9600, 19200, 38400 (19,200 default)
Data format: 8 data bits, no parity, one stop bit
Flow Control: XON/XOFF
Echo: off
Common Mode Voltage
Precision Analog Input: 150 (300 on channels 1 and 11) V dc or
ac RMS
Fast Analog Input: 50 V dc or 30 V ac rms
Operating Temperature
Range
-20
o
C to 60
o
C (-4
o
F to +140
o
F)
Storage Temperature
Range
-40
o
C to 70
o
C (-40
o
F to +158
o
F)
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...