2680A/2686A
Users Manual
A-10
Table A-12. PAI Module AC Voltage Accuracy Specifications
1 Year Ac (%input + V) [1]
Range Frequency
18
°
C to 28
°
C -10
°
C to 60
°
C
Slow
Fast
Slow
Fast
300 mV
20 to 50 Hz
3%+.25 mV
6%+.5 mV
3.5%+.25 mV
7%+.5 mV
50 to 150 Hz
0.4%+.25 mV
1%+.5 mV
0.5%+.25 mV
1.5%+.5 mV
150 Hz to 10 kHz
0.3%+.25 mV
1%+.5 mV
0.4%+.25 mV
1.5%+.5 mV
10 kHz to 20 kHz
0.4%+.25 mV
1%+.5 mV
0.7%+.25 mV
1.5%+.5 mV
20 kHz to 50 kHz
2%+.3 mV
3%+.5 mV
3%+.3 mV
4%+.5 mV
50 kHz to 100 kHz
5%+.5 mV
5%+1 mV
7%+.5 mV
8%+1 mV
3 V
20 to 50 Hz
3%+2.5 mV
6%+5 mV
3.5%+2.5 mV
7%+5 mV
50 to 150 Hz
0.4%+2.5 mV
1%+5 mV
0.5%+2.5 mV
1.2%+5 mV
150 Hz to 10 kHz
0.3%+2.5 mV
1%+5 mV
0.4%+2.5 mV
1.2%+5 mV
10 kHz to 20 kHz
0.4%+2.5 mV
1%+5 mV
0.5%+2.5 mV
1.2%+5 mV
20 kHz to 50 kHz
1%+3 mV
1.5%+6 mV
1.5%+3 mV
2%+6 mV
50 kHz to 100 kHz
2%+5 mV
3%+10 mV
3%+5 mV
4%+10 mV
30 V
20 to 50 Hz
3%+25 mV
6%+50 mV
3.5%+25 mV
7%+50 mV
50 to 150 Hz
0.4%+25 mV
1%+50 mV
0.5%+25 mV
1.2%+40 mV
150 Hz to 10 kHz
0.3%+25 mV
1%+50 mV
0.5%+25 mV
1.2%+40 mV
10 kHz to 20 kHz
0.4%+25 mV
1%+50 mV
0.5%+25 mV
1.2%+40 mV
20 kHz to 50 kHz
1%+30 mV
1.5%+60 mV
1%+30 mV
2%+50 mV
50 kHz to 100 kHz, V<20V
2%+50 mV
3%+100 mV
2.5%+50 mV
4%+100 mV
150/300 V
20 to 50 Hz
3%+.25 V
6%+.5 V
3.5%+.25 V
7%+.5 V
50 to 150 Hz
0.4%+.25 V
1%+.5 V
0.5%+.25 V
1.2%+.4 V
150 Hz to 2 kHz
Vx Hz<2 x10
6
0.3%+.25 V
1.2%+.5 V
0.5%+.25 V
1.4%+.4 V
2 kHz to 20 kHz, V<100V
0.4%+.25 V
1.6%+.5 V
0.5%+.25 V
1.8%+.4 V
20 kHz to 50 kHz, V<40V
1%+.30 V
2.0%+.6 V
1.2%+.30 V
2.5%+.5 V
[1] Sinewave inputs>6% of scale and signals with crest factors <2.
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...