FAI Module 4-Wire Resistance Measurement Specifications
The following tables provide FAI Module specifications for the 4-wire resistance
measurement function. The 4-wire measurements use 2 input channels a decade
apart, e.g., channels 4 and 14.
Table A-28. FAI Module 4-Wire Resistance Temperature Coefficient
Specification Characteristic
Temperature
Coefficient
Add 1/10th the 90 day specification per
°
C above 28
°
C or below
18
°
C.
Accuracy at -20
°
C
Multiply the -10
°
C to +60
°
C accuracy specification by 2. After 1 hour
warm-up. For accuracy between -10
°
C and -20
°
C, interpolate
linearly.
Table A-29. FAI Module 4-Wire Resistance Range and Resolution Specifications
Resolution
Current
Full Scale
Maximum Voltage
Range
Slow Fast Applied
Voltage
Applied by Instrument
300
Ω
10
m
Ω
30
m
Ω
1 mA
300 mV
3.5 V
3 k
Ω
100
m
Ω
300
m
Ω
100
µ
A
300 mV
3.5 V
30 k
Ω
1
Ω
3
Ω
10
µ
A
300 mV
3.5 V
300 k
Ω
10
Ω
30
Ω
10
µ
A
3.0 V
3.5 V
3 M
Ω
100
Ω
300
Ω
1
µ
A
3.0 V
3.5 V