Appendices
Specifications
A
A-25
Table A-34. FAI Module Thermocouple Specifications
Accuracy
+
°
C
Thermocouple Resolution
18
°
C to 28
°
C -10
°
C to 60
°
C
ITS90
°
C
90 Day
1 Year
1 Year
Type Temperature
°
C
Slow Slow Fast Slow Fast
E
-100 to -25
.3
0.80
0.90
1.50
1.00
1.60
-25
to
20
.2
0.70 0.70 1.20 0.80 1.30
20
to
600
.2
0.60 0.70 1.10 0.80 1.20
600 to 1000
.2
0.60
0.80
1.20
1.10
1.50
J
-100 to 80
.3
0.80
0.90
1.60
0.90
1.70
80
to
230
.2
0.70 0.80 1.40 0.90 1.50
230 to 760
.2
0.70
0.80
1.30
1.00
1.50
K
-100 to -25
.4
1.00
1.10
2.00
1.20
2.10
-25
to
120
.3
0.80 0.90 1.70 1.00 1.80
120 to 1000
.3
0.90
1.10
1.80
1.50
2.20
1000 to 1372
.3
1.20
1.50
2.30
2.00
2.90
L
-100 to 100
.4
1.20
1.30
2.90
1.60
3.10
100 to 800
.4
0.90
1.00
2.10
1.20
2.30
800 to 900
.3
0.70
0.80
1.30
1.00
1.50
N
-100 to -25
.5
1.40
1.50
2.80
1.50
2.90
-25
to
120
.5
1.10 1.30 2.30 1.30 2.40
120 to 1000
.4
1.00
1.10
2.00
1.20
2.10
1000 to 1300
.3
1.00
1.20
1.90
1.60
2.40
R
250 to 600
1
2.40
2.70
5.60
2.80
5.70
600 to 1500
1
2.00
2.30
4.60
2.40
4.80
1500 to 1767
1
2.00
2.30
4.50
2.80
5.10
S
250 to 1000
1
2.60
2.80
5.90
2.90
6.00
1000 to 1400
1
2.00
2.30
4.60
2.60
5.00
1400 to 1767
1
2.30
2.70
5.30
3.30
5.90
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...