2680A/2686A
Users Manual
4-10
2.
Connect the 5700A to Channel 1 of the First Module
When testing each
module, connect a cable from the Output VA HI and LO connectors of the
5700A to the Universal Input module terminals for channel 1 be certain to
connect the 5700A HI to terminal H and LO to terminal L. Insert the Universal
Input module into the instrument under test (Figure 4-3).
Table 4-2. Recommended Test Equipment
Device Type
Minimum Specifications
Recommended Model
Multifunction
DC Voltage: Fluke
5700A
Calibrator
Range: 90 mV to 300 V dc
Accuracy: 0.002%
AC Voltage:
Frequency
Voltage
Accuracy
1 kHz
29 mV to 300 V
0.05%
100 kHz
15 mV to 300 V
0.5%
Frequency:
10 kHz
1 V rms
0.01%
Ohms:
Ohms
Accuracy
190
Ω
0.005%
1.9
k
Ω
0.005%
19
k
Ω
0.005%
190
k
Ω
0.005%
1.9
M
Ω
0.005%
Digital Multimeter
General Purpose Measurement
Fluke 77
Signal Generator
Sine wave. 0.5 to 1 V rms, 10 Hz to 5 kHz
Philips PM5193
Oscilloscope
Alternative Equipment List
Device Type
Recommended Model
Multiproduct Calibrator
Fluke 5520
Function/Signal Generator
Philips PM5139 or HP33120A or equivalent
Decade Resistance Source
General Resistance RDS 66A or equivalent
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...