Maintenance
Accuracy Performance Tests
4
4-19
3.
Verify Accuracy
Configure the 5700A for the output values below and verify
the measurement is between the minimum and maximum values. Change the
channel 1 range as required (see Step 1).
Volts AC Range
5700A Output
Minimum Reading
Maximum Reading
300 mV
20 mV@1 kHz
0.0197 V
0.0203 V
300 mV
20 mV@100 kHZ
0.0185 V
0.0215 V
300 mV
300 mV@1 kHz
0.29885 V
0.30115 V
300 mV
300 mV@100 kHz
0.2845 V
0.3155 V
3 V
3 V@1 kHz
2.9885 V
3.0115 V
30 V
30 V@1 kHz
29.885 V
30.115 V
300 V
300 V@1 kHz
298.85 V
301.15 V
Frequency Accuracy Test
The following procedure tests the accuracy of the frequency function for the PAI
and FAI modules. Measurement accuracy applies to all channels, not just the
channel used for the test.
1.
Configure Channel 1 for Frequency
Using the Fluke DAQ software,
configure channel 1 for frequency. There is no range selection for frequency as
all frequency measurements use auto-ranging.
2.
Open Spy
Open the Spy window under the Configuration dialog.
3.
Verify Accuracy
Configure the 5700A for the output values below and verify
the measurement is between the minimum and maximum values. Change the
channel 1 range as required (see Step 1).
Frequency Range
5700A Output
Minimum Reading
Maximum Reading
Autorange Only
1 V@10 kHz
9.994 kHz
10.006 kHz
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...